Evaluation of a LED-based flatbed document scanner for radiochromic film dosimetry in transmission mode
José Manuel Lárraga-Gutiérrez1, Olivia Amanda García-Garduño1, Carlos Treviño-Palacios2
1Laboratorio de Física Médica, Instituto Nacional de Neurología y Neurocirugía, Insurgentes sur 3877, La Fama, Tlalpan 14269, Cd. de México, Mexico.
Abstract:
Flatbed scanners are the most frequently used reading instrument for radiochromic film dosimetry because its low cost, high spatial resolution, among other advantages. These scanners use a fluorescent lamp and a CCD array as light source and detector, respectively. Recently, manufacturers of flatbed scanners replaced the fluorescent lamp by light emission diodes (LED) as a light source. The goal of this work is to evaluate the performance of a commercial flatbed scanner with LED based source light for radiochromic film dosimetry. Film read out consistency, response uniformity, film-scanner sensitivity, long term stability and total dose uncertainty was evaluated. In overall, the performance of the LED flatbed scanner is comparable to that of a cold cathode fluorescent lamp (CCFL). There are important spectral differences between LED and CCFL lamps that results in a higher sensitivity of the LED scanner in the green channel. Total dose uncertainty, film response reproducibility and long-term stability of LED scanner are slightly better than those of the CCFL. However, the LED based scanner has a strong non-uniform response, up to 9%, that must be adequately corrected for radiotherapy dosimetry QA. The differences in light emission spectra between LED and CCFL lamps and its potential impact on film-scanner sensitivity suggest that the design of a dedicated flat-bed scanner with LEDs may improve sensitivity and dose uncertainty in radiochromic film dosimetry.
Related Concept Videos
Role of Communication in the Nursing Process III: Evaluation and Documentation
What is a Mode?
There can be more than one mode in a data set if multiple values have the same highest frequency. For instance, suppose that the Statistics exam scores of 20 students are: 50; 53; 59; 59; 63; 63; 72; 72; 72; 72; 72; 76; 78; 81; 83; 84; 84; 84; 90; 93. Here, the mode is 72, as it occurs most frequently, five times.
A data set with two modes is called bimodal. For example,...
Nursing Process for Patient and Caregiver Teaching III: Evaluation and Documentation
Nurses can use several methods to evaluate patient outcomes. For example, oral questions can assess cognitive learning,...
Transmission-based Precautions II: Airborne and Protective Environment
Airborne precautions:
Use airborne precautions when treating patients known or suspected to have diseases that spread through the air—for example, tuberculosis or measles. These organisms are present in smaller droplets expelled by an infected person and...
Transmission-based Precautions I: Contact, Enteric, and Droplets
Contact Precautions:
Contact precautions are the measures taken to prevent the transmission of infectious agents, especially epidemiologically important microorganisms such as MRSA or influenza, primarily transmitted through direct or indirect contact with an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...