Motion Artifact Reduction From High-Pitch Dual-Source Computed Tomography Pulmonary Angiography
Paul M Bunch, Urvi P Fulwadhva, Jeremy R Wortman
1Siemens Medical Solutions USA, Inc, Malvern, PA.
High-pitch dual-source (DS) computed tomography pulmonary angiography (CTPA) significantly reduces motion artifacts compared to single-source (SS) CTPA. This advanced technique improves image quality and lowers radiation dose, benefiting patient care.
Area of Science:
- Radiology
- Medical Imaging
- Cardiovascular Imaging
Background:
- Motion artifacts degrade image quality in computed tomography pulmonary angiography (CTPA).
- Dual-source (DS) CT technology offers potential advantages over single-source (SS) CT for reducing motion artifacts.
Purpose of the Study:
- To compare quantitative and qualitative measures of motion artifact between high-pitch DS and SS CTPA protocols.
- To evaluate the impact of DS versus SS CTPA on aortic, cardiac, and respiratory motion artifact.
Main Methods:
- Retrospective review of 80 non-electrocardiogram-gated CTPA examinations (40 SS, 40 DS).
- Quantitative assessment of motion artifact using maximal excursion of "double images" for aortic, left ventricular, and diaphragmatic motion.
- Qualitative assessment of motion artifact by three independent readers; evaluation of contrast enhancement, image noise, and radiation dose.
Main Results:
- DS CTPA demonstrated significantly reduced motion artifacts compared to SS CTPA, both quantitatively and qualitatively (P ≤ 0.004).
- DS protocol showed increased pulmonary arterial contrast enhancement but a lower enhancement-to-noise ratio due to increased image noise.
- Mean radiation dose was 60% lower with the DS protocol.
Conclusions:
- High-pitch DS CTPA effectively minimizes motion artifacts from the ascending aorta, heart, and diaphragm.
- DS CTPA offers improved image quality and reduced radiation exposure compared to SS CTPA for pulmonary angiography.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
11:19Dorsal Column Steerability with Dual Parallel Leads using Dedicated Power Sources: A Computational Model
Published on: February 10, 2011
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Perceiving Loudness, Pitch, and Location
Place theory, or place coding, suggests that different pitches are heard because various sound waves activate specific locations along the cochlea's basilar membrane. The brain determines the pitch of a sound by...
Oxidation-Reduction Reactions
Equation of Motion: General Plane motion
Moreover, the body's center of mass experiences a rotational effect as a result of these couple moments. This rotation can be articulated as the...
Absolute Motion Analysis- General Plane Motion
As the drone's propellers rotate, an upward force is generated that counteracts the force of gravity, enabling the drone to lift off from the ground. This initial movement of the drone is along a straight path, representing a form of translational motion. In this phase, every point on the...