Related Experiment Video
Updated: Feb 12, 2026

Cardiopulmonary Bypass in a Mouse Model: A Novel Approach
Published on: September 22, 2017
A novel wall water system for cardiopulmonary bypass may reduce the risk of aerosolized infection
Gregory S Matte1, Thomas J Sandora2, Robert J Howe1
1Department of Cardiac Surgery, Boston Children's Hospital, Boston, Mass.
Objective:
Oxygenators for cardiopulmonary bypass require water flow for their integral heat exchanger. Heater-cooler units are nearly universally used for this requirement. Heater-cooler units pose the risk of aerosolized infection. The Centers for Disease Control and Prevention recommended discontinuing use of Stöckert 3T heater-cooler units (LivaNova PLC, London, United Kingdom) in October 2016 because of this risk. We aimed to reduce the risk of aerosolized infection posed by heater-cooler units by eliminating those devices from our operating rooms.
Methods:
The cardiac surgery division collaborated with in-house specialties to engineer a novel wall water system. The design called for service to 4 operating rooms with the actual water mixing valve in an operating room closet. Remote temperature control was mounted next to the heart-lung machine. Primary safety systems built into the water system include 5 μm filtration, pressure regulating and relief valves, flow quantifiers, limits to the hot and chilled input temperatures, and a novel bridge near the heart-lung machine that allows the perfusionist to test the system before patient use and to quickly disconnect the patient in case of system malfunction. In addition, all water line connections can be made with the tubing drained and never under pressure.
Results:
This novel wall water system has successfully provided heat exchanger water flow on 625 patients undergoing congenital heart surgery requiring cardiopulmonary bypass during its first 9 months of use.
Conclusions:
Wall water systems are an option for oxygenator heat exchangers that allow for improved heat exchange performance while reducing the risk of heater-cooler unit-associated infection during cardiac surgery.
More Related Videos
09:54A Recovery Cardiopulmonary Bypass Model Without Transfusion or Inotropic Agents in Rats
Published on: March 23, 2018
05:32Povidone Iodine Rectal Preparation at Time of Prostate Needle Biopsy is a Simple and Reproducible Means to Reduce Risk of Procedural Infection
Published on: September 21, 2015
Related Concept Videos
Factors Affecting the Risk of Infection
The integrity and count of the white blood cells help the body resist pathogens and fight infection. When impaired, it reduces the body's resistance to pathogens. The acidic pH levels of the gastrointestinal, genitourinary tracts, and skin...
Adaptations that Reduce Water Loss
Plant Cell Wall
Cardiopulmonary Resuscitation I: Adult
States of Water
Water freezes when the intermolecular forces are greater than the kinetic energy. Unlike most other substances, water is less dense in its solid state than in its liquid state. This is because each water molecule can form...
Relative Risk