Related Experiment Video
Updated: Feb 12, 2026

Taste Exam: A Brief and Validated Test
Published on: August 17, 2018
The Validation of the Sample6 DETECTTM HT/L for AOAC Research Institute
Kakolie Banerjee1, Brittney Peirson1, Chuxuan Hu1
1Sample6, 12 Gill St, Suite 4200, Woburn, MA 01801.
Background:
Listeria spp. are an important foodborne human pathogen because of their ability to cause disease and high mortality in individuals, particularly pregnant women, neonates, the elderly, immunocompromised individuals, and children. The Sample6 DETECTTM HT/L Kit is a semi-automated qualitative pathogen detection system designed to detect Listeria spp. (L. monocytogenes, L. innocua, L. ivanovii, L. seeligeri, L. welshimeri, and L. marthii) in environmental samples using the Sample6 BioIlluminationTM technology.
Objective:
The study was done to evaluate the Sample6 DETECT HT/L Kit. The assay was evaluated for inclusivity, exclusivity, robustness, product consistency, and stability, and a matrix study of one environmental surface.
Methods:
The performance of the Sample6 DETECT HT/L was compared with U.S. Food and Drug Administration reference culture method for Listeria using an unpaired study design.
Results:
The Sample6 DETECT HT/L assay correctly identified all 50 inclusivity isolates and correctly excluded all 30 nontarget strains evaluated. The assay was not affected by minor variations in incubation temperature and time, or sample volume. Results across three production lots spanning the shelf life of the assay were consistent. In the matrix study, the Sample6 DETECT HT/L for Listeria correctly identified each test portion for the presence or absence of Listeria, and there were no statistically significant differences between candidate and reference method results.
Conclusions:
The data collected in this study demonstrate that the Sample6 DETECT HT/L assay is a reliable method for the detection of Listeria spp. on stainless-steel environmental surfaces after 22 h of enrichment.
Related Concept Videos
Reliability and Validity
Data Validation
Key parameters for method validation include:
Data Validation
Nursing assessment guides are generally based on holistic models rather than medical...
Formulating and Validating Nursing Diagnosis I
There are thirteen domains...
Formulating and Validating Nursing Diagnosis II
Risk nursing diagnoses represent clinical judgments of an individual, family, or community more vulnerable to developing the health problem than others...
In Vitro Drug Release Testing: Overview, Development and Validation