Contact-Free Screening of Atrial Fibrillation by a Smartphone Using Facial Pulsatile Photoplethysmographic Signals
Bryan P Yan1, William H S Lai2, Christy K Y Chan2
1Division of Cardiology, Department of Medicine and Therapeutics, Faculty of Medicine, The Chinese University of Hong Kong and Prince of Wales Hospital, Hong Kong SAR, China bryan.yan@cuhk.edu.hk.
Background:
We aimed to evaluate a novel method of atrial fibrillation (AF) screening using an iPhone camera to detect and analyze photoplethysmographic signals from the face without physical contact by extracting subtle beat-to-beat variations of skin color that reflect the cardiac pulsatile signal.
Methods And Results:
Patients admitted to the cardiology ward of the hospital for clinical reasons were recruited. Simultaneous facial and fingertip photoplethysmographic measurements were obtained from 217 hospital inpatients (mean age, 70.3±13.9 years; 71.4% men) facing the front camera and with an index finger covering the back camera of 2 independent iPhones before a 12-lead ECG was recorded. Backdrop and background light intensity was monitored during signal acquisition. Three successive 20-second (total, 60 seconds) recordings were acquired per patient and analyzed for heart rate regularity by Cardiio Rhythm (Cardiio Inc, Cambridge, MA) smartphone application. Pulse irregularity in ≥1 photoplethysmographic readings or 3 uninterpretable photoplethysmographic readings were considered a positive AF screening result. AF was present on 12-lead ECG in 34.6% (n=75/217) patients. The Cardiio Rhythm facial photoplethysmographic application demonstrated high sensitivity (95%; 95% confidence interval, 87%-98%) and specificity (96%; 95% confidence interval, 91%-98%) in discriminating AF from sinus rhythm compared with 12-lead ECG. The positive and negative predictive values were 92% (95% confidence interval, 84%-96%) and 97% (95% confidence interval, 93%-99%), respectively.
Conclusions:
Detection of a facial photoplethysmographic signal to determine pulse irregularity attributable to AF is feasible. The Cardiio Rhythm smartphone application showed high sensitivity and specificity, with low negative likelihood ratio for AF from facial photoplethysmographic signals. The convenience of a contact-free approach is attractive for community screening and has the potential to be useful for distant AF screening.
Related Concept Videos
Contact-dependent Signaling
Gap Junctions
In animal cells, gap junctions are formed...
Amyloid Fibrils
Amyloid deposits were observed as early as 1639 in the liver and the spleen. In 1854, Rudolph Virchow performed iodine staining,...
Amyloid Fibrils
Fibril-associated Collagen
For example, the type II collagen fibrils in cartilage have covalently bound type IX fibril-associated collagens at regular intervals. Other types of fibril-associated collagens are...
Contact Angle
The adhesive force is the molecular force between molecules of different materials, that is, between the molecules of the solid and the liquid. The cohesive...
Muscles for Facial Expressions