Influence of soft tissue in the assessment of the primary fixation of acetabular cup implants using impact analyses
Romain Bosc1, Antoine Tijou2, Giuseppe Rosi2
1NSERM U955, Team 16, Créteil, France; Vaccine Research Institute (VRI), Faculté de Médecine, Créteil, France; Université Paris Est, Créteil, France; Hopital Henri Mondor, Plastic, Reconstructive, Aesthetic and Maxillofacial Surgery Department, 50, avenue du Maréchal de Lattre de Tassigny, 94000 Créteil, France.
Background:
The acetabular cup (AC) implant primary stability is an important determinant for the success of cementless hip surgery but it remains difficult to assess the AC implant fixation in the clinic. A method based on the analysis of the impact produced by an instrumented hammer on the ancillary has been developed by our group (Michel et al., 2016a). However, the soft tissue thickness present around the acetabulum may affect the impact response, which may hamper the robustness of the method. The aim of this study is to evaluate the influence of the soft tissue thickness (STT) on the acetabular cup implant primary fixation evaluation using impact analyses.
Methods:
To do so, different AC implants were inserted in five bovine bone samples. For each sample, different stability conditions were obtained by changing the cavity diameter. For each configuration, the AC implant was impacted 25 times with 10 and 30 mm of soft tissues positioned underneath the sample. The averaged indicator Im was determined based on the amplitude of the signal for each configuration and each STT and the pull-out force was measured.
Findings:
The results show that the resonance frequency of the system increases when the value of the soft tissue thickness decreases. Moreover, an ANOVA analysis shows that there was no significant effect of the value of soft tissue thickness on the values of the indicator Im (F = 2.33; p-value = 0.13).
Interpretation:
This study shows that soft tissue thickness does not appear to alter the prediction of the acetabular cup implant primary fixation obtained using the impact analysis approach, opening the path towards future clinical trials.
Related Concept Videos
Impact of Groups on Groups
Fixation and Sectioning
The simplest type of preparation is the wet mount, in which the specimen is placed in a drop of liquid on the slide. A liquid specimen can be directly deposited on the slide using a dropper. Solid specimens, such as skin scraping, can be placed on the slide before adding a drop of liquid to prepare the wet mount. Sometimes the liquid is simply water, but stains are often added...
Impact
When particles with different initial velocities collide, they induce deformation by applying equal and opposite impulses. At the point of maximum deformation, the particles move together with...
Carbon-dioxide Fixation
Impact of Groups on Individuals
Impact of Individuals on a Group