Likely False-Positive Pneumococcal Antigen Test BinaxNOW Due to Parvimonas micra: A Four-Case Series
Marie-Caroline Ploton1, Marion Caseris1, Christelle Jost2
1AP-HP, Equipe opérationnelle d'Infectiologie, Hôpital Robert-Debré, Paris, France.
Abstract:
We retrospectively report four cases from two hospitals of nonpneumococcal pleural empyema with a likely false-positive result on the pneumococcal antigen test BinaxNOW (PATB) (Alere) performed in pleural fluid samples in patients with aspiration pneumonia risk factors. To determine whether the positive reaction was due to cross-reactivity, we separately tested the isolates from the pleural fluid samples, along with collection and reference strains. All patients had polymicrobial aerobic and anaerobic positive cultures, including Parvimonas micra in every case. In all cases, 16S rDNA polymerase chain reaction sequencing yielded Fusobacterium nucleatum. Samples for culture and specific polymerase chain reaction were negative for Streptococcus pneumoniae. We found that the false-positive PATB finding was likely due to P micra, a previously unknown cross-reactivity. In case of aspiration pneumonia risk factors, a positive PATB result must be interpreted with caution because there can be a false positivity due to anaerobic infection or co-infection.
Insights
A common test for pneumococcal pneumonia may show false positives in patients with aspiration pneumonia risk factors. This cross-reactivity is likely due to anaerobic bacteria, not Streptococcus pneumoniae.
Area of Science:
- Microbiology
- Infectious Diseases
- Clinical Diagnostics
Background:
- Pleural empyema can be challenging to diagnose, especially in patients with risk factors for aspiration pneumonia.
- The BinaxNOW pneumococcal antigen test (PATB) is a rapid diagnostic tool used to detect Streptococcus pneumoniae in pleural fluid.
- Accurate diagnosis is crucial for appropriate antibiotic selection and patient management.
Observation:
- Four cases of nonpneumococcal pleural empyema were identified in patients with aspiration pneumonia risk factors.
- These patients exhibited false-positive results on the pneumococcal antigen test (PATB) performed on pleural fluid.
- All patients presented with polymicrobial infections, including Parvimonas micra and Fusobacterium nucleatum.
Findings:
- Cultures and PCR testing were negative for Streptococcus pneumoniae in all cases.
- Isolates from pleural fluid, including Parvimonas micra, were tested to investigate cross-reactivity with the PATB.
- The study identified a previously unrecognized cross-reactivity of the PATB with Parvimonas micra, leading to false-positive results.
Implications:
- Positive PATB results in patients with aspiration pneumonia risk factors should be interpreted with caution.
- The presence of anaerobic bacteria, such as Parvimonas micra, can cause false-positive PATB results.
- This finding necessitates a re-evaluation of diagnostic strategies for pleural empyema in at-risk populations.
More Related Videos
Related Concept Videos
False Memories
One primary source of false memories is misattribution, where individuals incorrectly associate external information...
Resistors In Series
In a series circuit, the...
Series Resonance
Position-effect Variegation
Series and Parallel Capacitors
First, consider capacitors connected in series to a battery. In this configuration, the plate connected to the battery's positive terminal develops a positive charge, while the plate attached to the negative terminal becomes negatively charged. An equal magnitude of charge is induced on the...
Trigonometric Fourier series
The trigonometric Fourier series specifically expresses a periodic function with a defined period T using sine...