Comparison of Keratometry Obtained by a Swept Source OCT-Based Biometer with a Standard Optical Biometer and
Leyla Asena1, Ahmet Akman1, Sirel Gür Güngör1
1a Department of Ophthalmology , Başkent University Faculty of Medicine , Ankara , Turkey.
Purpose:
To assess agreement of a swept source-optical coherence tomography (SS-OCT) based Biometer with a standard IOLMaster device and Scheimpflug Imaging (SI) to acquire keratometric measurements in cataract patients.
Methods:
In this prospective comparative study, 101 eyes of 101 cataract surgery candidates, aged 24-81 years, were sequentially examined using three devices. Keratometry values at the flat (K1) and steep (K2) axis, mean corneal power (Km) and magnitude of corneal astigmatism as well as J0 and J45 vectoral components of astigmatism obtained with the SS-OCT based biometer (IOLMaster 700) were compared with those obtained with the IOLMaster 500 and SI. The agreement between measurements was evaluated by the Bland-Altman method, intraclass correlation coefficients (ICCs) and repeated-measures analysis of variance.
Results:
Mean K1 values from the three devices were similar (p = 0.09). Mean K2 and Km values of IOLMaster 700 were higher than SI and lower than IOLMaster 500 (p = 0.04 for K2 and p = 0.02 for Km). There was a strong correlation between K1, K2, Km and magnitude of astigmatism obtained with all devices (r ≥ 0.80 and p < 0.01). The 95% limits of agreement (LoA) width for each keratometric value were highest for the comparison between IOLMaster 500 and SI and lowest for the comparison between IOLMaster 700 and 500. The mean differences (width of 95% LoA) for J0 and J45 vectoral components were 0.005 (2.19) and 0.12 (2.92) for the measurements obtained by IOLMaster 700 vs IOLMaster 500 and 0.06 (1.79) and 0.02 (1.58) for the measurements obtained by IOLMaster 700 vs SI, respectively.
Conclusions:
With ICCs close to 1, the agreement between all devices was excellent for keratometric measurements. Mean K2, Km and astigmatism measurements from IOLMaster 700 were lower than IOLMaster 500 and higher than SI. However, the differences were quite small and are not expected to affect the final IOL power.
More Related Videos
12:22Multimodal Volumetric Retinal Imaging by Oblique Scanning Laser Ophthalmoscopy oSLO and Optical Coherence Tomography OCT
Published on: August 4, 2018
08:13A Guide to Build a Highly Inclined Swept Tile Microscope for Extended Field-of-view Single-molecule Imaging
Published on: April 8, 2019
Related Concept Videos
Sources of Self-Esteem III: Social Comparison
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Measurement: Standard Units
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Standard Enthalpy of Formation