Related Experiment Video
Updated: Feb 12, 2026

C-arm-Free Simultaneous OLIF51 and Percutaneous Pedicle Screw Fixation in a Single Lateral Position
Published on: September 16, 2022
An intraoperative fluoroscopic method to accurately measure the post-implantation position of pedicle screws
Robyn Newell1,2, Hooman Esfandiari3,4, Carolyn Anglin5,6
1Biomedical Engineering, University of British Columbia, Vancouver, BC, Canada. robyn.newell@gmail.com.
Purpose:
Pedicle screw malplacement, leading to neurological symptoms, vascular injury, and premature implant loosening, is not uncommon and difficult to reliably detect intraoperatively with current techniques. We propose a new intraoperative post-placement pedicle screw position assessment system that can therefore allow surgeons to correct breaches during the procedure. Our objectives were to assess the accuracy and robustness of this proposed screw location system and to compare its performance to that of 2D planar radiography.
Methods:
The proposed system uses two intraoperative X-ray shots acquired with a standard fluoroscopic C-arm and processed using 2D/3D registration methods to provide a 3D visualization of the vertebra and screw superimposed on one another. Point digitization and CT imaging of the residual screw tunnel were used to assess accuracy in five synthetic lumbar vertebral models (10 screws in total). Additionally, the accuracy was evaluated with and without correcting for image distortion and for various screw lengths, screw materials, breach directions, and vertebral levels.
Results:
The proposed method is capable of localizing the implanted screws with less than 2 mm of translational error (RMSE: 0.7 and 0.8 mm for the screw head and tip, respectively) and less than [Formula: see text] angular error (RMSE: [Formula: see text]), with minimal change to the errors if image distortion is not corrected. Breaches and their anatomical locations were all correctly visualized and identified for a variety of screw lengths, screw materials, breach locations, and vertebral levels, demonstrating the robustness of this approach. In contrast, one breach, one non-breach, and the anatomical location of three screws were misclassified with 2D X-ray.
Conclusion:
We have demonstrated an accurate and low-radiation technique for localizing pedicle screws post-implantation that requires only two X-rays. This intraoperative feedback of screw location and direction may allow the surgeon to correct malplaced screws intraoperatively, thereby reducing postoperative complications and reoperation rates.
Related Concept Videos
Velocity and Position by Integral Method
Consider an example to calculate the velocity and position from the acceleration function. A motorboat is traveling at a constant velocity of 5.0 m/s when it starts to decelerate to arrive at the dock. Its acceleration is...
Screw: Problem Solving
To evaluate the...
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Velocity and Position by Graphical Method
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
Position-effect Variegation