Evaluation of Left Ventricular Structure and Function using 3D Echocardiography
Published on: October 28, 2020
What is the minimum change in left ventricular ejection fraction, which can be measured with contrast
Tan Suwatanaviroj1, Weimin He1, Edith Pituskin2
1Mazankowski Alberta Heart Institute, University of Alberta Hospital, Edmonton, Canada.
Background:
There are limited data on what is the minimum change that can be detected in cancer patients undergoing treatment with cardiotoxic drugs and are referred for monitoring left ventricular (LV) function.
Objective:
To assess the variability in the measurement of LV volumes and ejection fraction (EF) in contrast echocardiography and to determine the minimum detectable difference (MDD) between two EF measurements that can be deemed significant.
Methods:
A total of 150 patients were divided into three groups according to EF (EF <53, 53-60, and >60%). Each group consisted of 50 randomly selected cancer patients who underwent contrast echocardiography between July 2010 and May 2014. Repeated measurements of LV volumes and EF were performed offline by a sonographer and a cardiologist. Inter-observer variability was assessed by analysis of variance. Measurement error was estimated by standard error of measurement and MDD.
Results:
The 95% confidence interval with a single measurement of EF was 2 percentage points in the groups of patients with EF <53% and EF >60%, and 2.5 percentage points for patients with EF 53-60%. The MDD for EF, end-diastolic volume and end-systolic volume that could be recognized with 95% confidence interval were 4 percentage points, 7 mL and 4 mL, respectively.
Conclusion:
Contrast echocardiography is a reliable tool for serial measurements of EF to monitor cardiotoxicity due to chemotherapy. In a high-volume echocardiography laboratory with experienced staff, the MDD for EF of 4 percentage points on a good-quality recording demonstrates the high reproducibility of the Simpson's method using contrast echocardiography.
More Related Videos
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Standard Entropy Change for a Reaction
Le Chatelier's Principle: Changing Concentration
Global Climate Change
Rates of Change
Work Done During Volume Change
Consider a gas confined to a cylinder fitted with a movable piston at one end. If the gas expands from volume V1 to volume V2, it exerts a force on the piston, such that the piston moves by a distance dr.
The work done by the gas on the piston can be expressed as