Radiation dose reduction in perfusion CT imaging of the brain using a 256-slice CT: 80 mAs versus 160 mAs
Isabelle Riederer1, Claus Zimmer2, Daniela Pfeiffer3
1Department of Diagnostic and Interventional Radiology, Klinikum rechts der Isar, Technische Universität München, Munich, Germany; Department of Diagnostic and Interventional Neuroradiology, Klinikum rechts der Isar, Technische Universität München, Munich, Germany.
Purpose:
To examine CTP of the brain in real patient data after reducing tube current down to 80 mAs to decrease radiation dose.
Methods:
CTP was acquired in 60 suspected stroke patients with 80 (n: 30) or 160 (n: 30) mAs. Data were analyzed retrospectively by two independent readers. SNR, perfusion maps and image quality were compared in hypoperfused and non-affected areas.
Results:
SNR was significantly higher in CTP with 160 mAs compared to 80 mAs (p < 0.001) in non-affected regions, but there was no significant difference in hypoperfused regions. Overall, images with 80 mAs were rated worse than the ones with 160 mAs (3.0 ± 0.7 versus 4.0 ± 0.7), however, still as sufficient to detect proximal vessel occlusions.
Conclusion:
Tube current of 80 mAs is still sufficient for the detection of perfusion deficits of proximal vessel occlusions.
More Related Videos
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Radiological Investigation I: X-ray and CT
Biological Effects of Radiation
Oxidation-Reduction Reactions