Gamma spectrum and coincidence summation simulations with Geant4 in the analysis of radionuclide using BEGe detector
Chuanlei Liu1, Weihua Zhang1, Dylan Pierce2
1Radiation Protection Bureau of Health Canada, Ottawa, ON, Canada K1A 1C1.
Abstract:
A Geant4 application has been developed to simulate the gamma spectrum of a BEGe detector. The summation effects were encompassed for both random and true coincidences through a time-based approach. The application well reproduces the typical spectral features such as full energy peaks, escape peaks, summation peaks and Compton/backscatter continuum. The simulated gamma response is consistent with empirical data within 3%, while simulations of the X-ray and baseline regions are valid within a 10% deviation.