Evaluation of low-intensity anti-coagulation with a fully magnetically levitated centrifugal-flow circulatory
Ivan Netuka1, Peter Ivák2, Zuzana Tučanová1
1Department of Cardiovascular Surgery, Institute for Clinical and Experimental Medicine, Prague, Czech Republic.
Background:
The HeartMate 3 left ventricular assist system is engineered to avoid pump thrombosis, yet bleeding complications persist. We investigated the safety of low-intensity anti-coagulation in patients with the HeartMate 3.
Methods:
The Minimal AnticoaGulation EvaluatioNTo aUgment heMocompatibility (MAGENTUM 1) pilot study is a prospective, single-arm study of low-intensity warfarin anti-coagulation in patients implanted with the HeartMate 3 pump. After standard warfarin anti-coagulation (international normalized ratio [INR] 2.0 to 3.0) and aspirin for 6 weeks post-implant, patients were transitioned to a lower INR target range of 1.5 to 1.9. The primary end-point was a composite of survival free of pump thrombosis, disabling stroke (modified Rankin score [MRS] >3), or major bleeding (excluding peri-operative bleeding) with at least 6-month post-implant follow-up. Time in therapeutic range (TTR) was measured to assess anti-coagulation target efficacy using the Rosendaal method. A safety algorithm to monitor for signs of pump thrombosis was developed and implemented.
Results:
We enrolled 15 patients (mean age 57.3 ± 13.3 years), 13 men with advanced heart failure (67% with INTERMACS Profiles 2 or 3), irrespective of therapeutic goal of bridge-to-transplant or destination therapy. The primary end-point was met in 14 of 15 (93 ± 6%) patients; 1 patient developed recurrent gastrointestinal bleeding. The TTR during the reduced anti-coagulation phase (6 weeks to 6 months) was 75.3 ± 8.6%. No thrombotic events occurred.
Conclusions:
This pilot study suggests low-intensity anti-coagulation targeting an INR between 1.5 and 1.9 is achievable and safe with the HeartMate 3 cardiac pump in the short-term phase, 6-months post-implant. A large-scale trial is now warranted.
Related Concept Videos
Anatomy of the Circulatory System
Coagulation
Coagulation
During the coagulation phase, clotting factors, or procoagulants, play a vital role in initiating and progressing the coagulation cascade. This cascade is a series of reactions...
Centrifugation
ATP Driven Pumps III: V-type Pumps
The peripheral or cytosolic V1 domain with eight subunits is involved in ATP hydrolysis. The integral or transmembrane V0 domain containing at least five subunits...
ATP Driven Pumps II: P-type Pumps
A typical P-type pump has three cytosolic domains: nucleotide-binding (N), phosphorylation (P), and activator (A) domains. These domains are connected to the membrane-spanning helices by short amino acid segments. ATP hydrolysis and covalent phosphoenzyme intermediate formation are crucial parts of the catalytic cycle. At the highly...