The robustness of T2 value as a trabecular structural index at multiple spatial resolutions of 7 Tesla MRI
D K Lee1, Y K Song2, B W Park1
1Department of Biomedical Engineering, Ulsan National Institute of Science and Technology, Ulsan, South Korea.
Purpose:
To evaluate the robustness of MR transverse relaxation times of trabecular bone from spin-echo and gradient-echo acquisitions at multiple spatial resolutions of 7 T.
Methods:
The effects of MRI resolutions to T2 and T2* of trabecular bone were numerically evaluated by Monte Carlo simulations. T2 , T2*, and trabecular structural indices from multislice multi-echo and UTE acquisitions were measured in defatted human distal femoral condyles on a 7 T scanner. Reference structural indices were extracted from high-resolution microcomputed tomography images. For bovine knee trabecular samples with intact bone marrow, T2 and T2* were measured by degrading spatial resolutions on a 7 T system.
Results:
In the defatted trabecular experiment, both T2 and T2* values showed strong ( |r| > 0.80) correlations with trabecular spacing and number, at a high spatial resolution of 125 µm3 . The correlations for MR image-segmentation-derived structural indices were significantly degraded ( |r| < 0.50) at spatial resolutions of 250 and 500 µm3 . The correlations for T2* rapidly dropped ( |r| < 0.50) at a spatial resolution of 500 µm3 , whereas those for T2 remained consistently high ( |r| > 0.85). In the bovine trabecular experiments with intact marrow, low-resolution (approximately 1 mm3 , 2 minutes) T2 values did not shorten ( |r| > 0.95 with respect to approximately 0.4 mm3 , 11 minutes) and maintained consistent correlations ( |r| > 0.70) with respect to trabecular spacing (turbo spin echo, 22.5 minutes).
Conclusion:
T2 measurements of trabeculae at 7 T are robust with degrading spatial resolution and may be preferable in assessing trabecular spacing index with reduced scan time, when high-resolution 3D micro-MRI is difficult to obtain.
Related Concept Videos
Structures of Solids
Covalent Bonding and Lewis Structures
Multiple Allele Traits
Multiple Regression
Farmers can use multiple regression to determine the crop yield based on more than one factor, such as water availability, fertilizer, soil properties, etc. Here, the crop yield is the response or dependent variable as it depends on the other independent variables. The analysis requires the construction of a scatter plot...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Depth Perception and Spatial Vision