Single-use disposable digital flexible ureteroscopes: an ex vivo assessment and cost analysis
Derek B Hennessey1, Grzegorz L Fojecki2, Nathan P Papa1
1Department of Urology, Austin Health, Melbourne, Vic., Australia.
Objective:
To assess and measure the capability of a single-use disposable digital flexible ureteroscope, the LithoVue™ (Boston Scientific, Marlborough, MA, USA), and to assess if there is a benefit to switching to single-use scopes.
Patients And Methods:
The LithoVue was compared to two commonly used reusable flexible ureteroscopes (Olympus URF-V [Olympus, Tokyo, Japan] and Karl Storz Flex-Xc [Karl Storz & Co. KG, Tuttlingen, Germany]) ex vivo. An analysis of reusable ureteroscope usage was performed to evaluate damage, durability, and maintenance costs. This was then compared to the projected costs of using single-use disposable scopes.
Results:
Flexion, deflection and irrigation flow of the LithoVue was equivalent, if not better than the reusable flexible ureteroscopes. An analysis of 234 procedures with seven new Olympus URF-V scopes, revealed 15 scope damages. Staghorn stones and lower pole/mid-zone stones were significant risk factors for damage (P = 0.014). Once damage occurred it was likely to occur again. Total repair costs were $162 628 (Australian dollars) (£92 411 in Great British pounds), the mean cost per case was $695 (£395). Factoring in the purchase cost, cleaning and repair costs, the cumulative cost of 28 reusable flexible ureteroscopy procedures was ~$50 000 (£28 412). If the LithoVue was priced at $1 200 (£682), switching to a single-use scope would cost ~$35 000 (£19 888).
Conclusion:
The LithoVue is analogous to reusable flexible ureteroscopes in regard to standard technical metrics. Depending on its purchase cost it may also represent a cost saving for hospitals when compared to the cumulative costs of maintaining reusable scopes. Additionally, urologist may consider using the scope in cases in which reusable scope damage is anticipated.
Related Concept Videos
¹H NMR of Conformationally Flexible Molecules: Temporal Resolution
Social Foundations of Self IV: Self in Digital Communication
Pedigree Analysis
¹H NMR of Conformationally Flexible Molecules: Variable-Temperature NMR
Dimensional Analysis
Conversion Factors and Dimensional Analysis
The unit...
Load along a Single Axis
Consider a beam of length L subjected to a varying load, which is a combination of parabolic and trapezoidal load distribution along the x-axis. In this case, it is essential to determine the resultant loads, their locations, and...