Related Experiment Video
Updated: Feb 11, 2026

The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report
Published on: August 4, 2022
High Rate of Early Revision After Custom-Made Unicondylar Knee Arthroplasty
Carl T Talmo1, Marie C Anderson2, Eliot S Jia2
1Department of Orthopaedic Surgery, New England Baptist Hospital, Boston, Massachusetts.
Background:
There has been a recent interest in custom-made partial knee arthroplasties to provide patient-specific instrumentation and better fit of the prosthesis. While unicondylar knee arthroplasties (UKAs) have demonstrated good outcomes and durable results in many studies, there is little evidence on outcomes of these custom-made implants.
Methods:
We performed a retrospective review of all custom-made UKAs performed at our institution by one surgeon from 2008 to 2015. We analyzed preoperative demographics, clinical follow-up evaluations, and radiographs and performed an analysis of risk factors including age, gender, height, weight, body mass index, and tibial insert thickness. The incidence of revision surgery, radiographic failures indicating component loosening, and symptomatic clinically failed implants was calculated at an average of 54.0 months of follow-up.
Results:
We analyzed 115 consecutive custom-made medial UKAs from a single surgeon at our institution and found 29 (25.2%) UKAs had failed at an average of 33.1 months after surgery. Reasons for failure included aseptic femoral loosening (10), aseptic tibial loosening (8), loosening of both components (4), infection (3), progression of osteoarthritis (2), pain (1), and dislodged polyethylene insert (1). We found a significant relationship between implant failure and body mass index; no other study variables were statistically significant.
Conclusion:
We found a relatively high rate of aseptic loosening and particularly femoral component loosening in the short- to intermediate-term follow-up period. While further study of larger numbers of custom-made UKA from multiple institutions may help verify these findings, we recommend careful consideration of the use of this implant.
More Related Videos
Related Concept Videos
Knee Joint
A total of seven ligaments support the knee joint. The patellar ligament, which is also attached to the quadriceps femoris...
Reaction Rate
The mathematical representation of the change in the concentration of reactants and products, over time, is the rate...
Related Rates
Speciation Rates
Measuring Reaction Rates
Rate-Determining Steps
In a multistep reaction mechanism, one of the elementary steps progresses significantly slower than the others. This slowest step is called the rate-limiting step (or rate-determining step). A reaction cannot proceed faster than its slowest step, and hence, the rate-determining step limits the overall reaction rate.
The concept of rate-determining step can be understood from the analogy of a 4-lane freeway with a short-stretch of traffic-bottleneck caused due to...