Enhancing multi-spot structured illumination microscopy with fluorescence difference

Edward N Ward1, Frida H Torkelsen1, Robert Pal1

  • 1Department of Chemistry, Durham University, South Road, Durham DH1 3LE, UK.

Related Concept Videos

Confocal Fluorescence Microscopy01:16

Confocal Fluorescence Microscopy

Confocal microscopy is an advanced microscopic technique. The prime advantage of the confocal microscope over other microscopy techniques is its ability to block the out-of-focus light from the illuminated samples using pinholes. It is widely used with fluorescence optics to obtain high-resolution, sharp contrast images. Unlike optical microscopes, confocal microscopes use a focused beam of light laser to scan the entire sample surface at different z-planes. These microscopes are, therefore,...
21.2K
Super-resolution Fluorescence Microscopy01:37

Super-resolution Fluorescence Microscopy

Super-resolution fluorescence microscopy (SRFM) provides a better resolution than conventional fluorescence microscopy by reducing the point spread function (PSF). PSF is the light intensity distribution from a point that causes it to appear blurred. Due to PSF, each fluorescing point appears bigger than its actual size, and it is the PSF interference of nearby fluorophores that causes the blurred image. Various approaches to achieving higher resolution through SRFM have recently been...
14.6K
Electric Potential and Potential Difference01:16

Electric Potential and Potential Difference

Suppose a positive test charge moves away from a positive static charge, then the Coulomb force does positive work, and its electric potential energy decreases. The potential energy per unit charge is defined as the electric potential. The electric potential is independent of the test charge.
When a test charge moves from the initial to the final position, the electric potential difference between those positions is defined as the ratio of the change in the potential energy to the charge on the...
5.8K
Difference from Background: Limit of Detection01:05

Difference from Background: Limit of Detection

The limit of detection (LOD) is the smallest amount of analyte that can be distinguished from the background noise. The LOD value corresponds to the concentration at which the analyte signal is three times larger than the standard deviation of the blank signal. Below this value, the analyte signal cannot be differentiated from the background noise. It is calculated by dividing the calibration slope by 3 times the standard deviation of the blank signals.
The LOD indicates the presence or absence...
8.4K
Total Internal Reflection Fluorescence Microscopy01:05

Total Internal Reflection Fluorescence Microscopy

Total internal reflection fluorescence microscopy or TIRF is an advanced microscopic technique used to visualize fluorophores in samples close to a solid surface with a higher refractive index, such as a glass coverslip. TIRF only allows fluorophores in proximity to the solid surface to be excited. When light from a medium with a lower refractive index (such as air) hits the glass coverslip at a critical angle, the light undergoes total internal reflection stead of passing through the glass.
13.5K
Identifying Statistically Significant Differences: The F-Test01:14

Identifying Statistically Significant Differences: The F-Test

The F-test is used to compare two sample variances to each other or compare the sample variance to the population variance. It is used to decide whether an indeterminate error can explain the difference in their values. The underlying assumptions that allow the use of the F-test include the data set or sets are normally distributed, and the data sets are independent of each other. The test statistic F is calculated by dividing one variance by another. In other words, the square of one standard...
3.9K