Quantitative determination of residual 1,4-dioxane in three-dimensional printed bone scaffold
Ling Li1, Jing Long1, Long Li1
1Translational Medicine R&D Center, Institute of Biomedical and Health Engineering, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen, China.
Background/Objective:
A novel porous scaffold poly (lactide-co-glycolide) and tricalcium phosphate (PLGA/TCP) was developed by three-dimensional printing technology for bone defect repair. As a Class 2 solvent with less severe toxicity, content of residual 1,4-dioxane in this newly developed scaffold should be rigorously controlled when it is translated to clinical use. In this study, a headspace gas chromatography-mass spectrometric (HS-GC-MS) method and related testing protocol were developed for quantitative determination of 1,4-dioxane in the PLGA/TCP composite scaffolds.
Methods:
Matrix effect analysis was used to optimise the pretreatment method of the scaffolds. Then, the procedure for testing 1,4-dioxane using HS-GC-MS was set up. The accuracy, precision, and robustness of this newly developed quantitative method were also validated before quantification of 1,4-dioxane in the scaffolds with different drying procedures.
Results:
Dimethyl formamide (DMF) was the optimal solvent for dissolving scaffolds for GC-MS with proper sensitivity and without matrix effect. Then, the optimised procedure was determined as: the scaffolds were dissolved in DMF and kept at 90°C for 40 minutes, separated on a HP-5MS column, and detected by mass spectroscopy. Recovery experiments gave 97.9-100.7% recovery for 1,4-dioxane. The linear range for 1,4-dioxane was determined as 1-40 ppm with linear correlation coefficient ≥ 0.9999. Intraday and interday precision was determined as being within relative standard deviation of below 0.68%. The passable drying procedure was related to lyophilising (-50°C, 50 Pa) the scaffolds for 2 days and drying in vacuum (50 Pa) for 7 days.
Conclusion:
This is the first quantitative method established to test 1,4-dixoane in a novel scaffold. This method was validated with good accuracy and reproducibility, and met the methodological requirements of the Guideline 9101 documented in the Chinese Pharmacopoeia 2015 Edition.
The Translational Potential Of This Article:
This quantitative method for determination of residual 1,4-dioxane in the novel scaffolds is a key technical method during its translation into clinical use because this method is an important and indispensable file in the enterprise standard when the porous scaffold is registered as a Class III implanted medical device for bone defect repair, which is used to guarantee the safety of the scaffolds. It is also applied to optimise the drying process of scaffolds and to monitor the quality of scaffolds in the industrialisation process. Further, this method provides references for other solvents quantitative determination in porous scaffolds or materials.
More Related Videos
09:34Visualizing Angiogenesis by Multiphoton Microscopy In Vivo in Genetically Modified 3D-PLGA/nHAp Scaffold for Calvarial Critical Bone Defect Repair
Published on: September 7, 2017
05:52Core/shell Printing Scaffolds For Tissue Engineering Of Tubular Structures
Published on: September 27, 2019
Related Concept Videos
Residual Plots
When the residual values are plotted against the variable x, it is called a residual...
Residual Stresses
Residual Stresses in Circular Shafts
Residual Stresses in Bending
Residuals and Least-Squares Property
If the observed data point lies above the line, the residual is positive, and the line underestimates the actual data value for y. If the observed data point lies below the line, the residual is negative, and the line overestimates the actual data value for y.
The process of fitting the best-fit...
Benzene to 1,4-Cyclohexadiene: Birch Reduction Mechanism