Related Experiment Video
Updated: Feb 11, 2026

Optimizing Minimally Invasive Spine Surgery: A Fully 3D CT O-Arm Navigated Workflow in MIS TLIF
Published on: October 17, 2025
Accuracy of image-guided percutaneous injection into a phantom spinal cord utilizing flat panel detector CT with MR
Jason F Talbott1,2, Daniel L Cooke1, Marc C Mabray3
1Department of Radiology and Biomedical Imaging, University of California San Francisco and Zuckerberg San Francisco General Hospital, San Francisco, California, USA.
Purpose:
To evaluate the accuracy of percutaneous fluoroscopic injection into the spinal cord of a spine phantom utilizing integrated navigational guidance from fused flat panel detector CT (FDCT) and MR datasets. Conventional and convection-enhanced delivery (CED) techniques were evaluated.
Materials And Methods:
FDCT and MR datasets of a swine thoracic spine phantom were co-registered using an integrated guidance system and surface to spinal cord target trajectory planning was performed on the fused images. Under real-time fluoroscopic guidance with pre-planned trajectory overlay, spinal cord targets were accessed via a coaxial technique. Final needle tip position was compared with a pre-determined target on 10 independent passes. In a subset of cases, contrast was injected into the central spinal cord with a 25G spinal needle or customized 200 µm inner diameter step design cannula for CED.
Results:
Average needle tip deviation from target measured 0.92±0.5 mm in the transverse, 0.47±0.4 mm in the anterior-posterior, and 1.67±1.2 mm in the craniocaudal dimension for an absolute distance error of 2.12±1.12 mm. CED resulted in elliptical intramedullary diffusion of contrast compared with primary reflux observed with standard needle injection.
Conclusions:
These phantom feasibility data demonstrate a minimally invasive percutaneous approach for targeted injection into the spinal cord utilizing real-time fluoroscopy aided by overlay trajectories derived from fused MRI and FDCT data sets with a target error of 2.1 mm. Intramedullary diffusion of injectate in the spinal cord is facilitated with CED compared with standard injection technique. Pre-clinical studies in large animal models are warranted.
More Related Videos
Related Concept Videos
Spinal Cord
The Spinal Cord
Spinal Cord: Information Processing
Sensory Information Processing
Sensory information processing begins at the sensory receptors located in the skin and other tissues, which detect somatic sensory stimuli such as touch, temperature, or pain. These receptors function as catalysts, initiating...
Spinal Cord: Gross Anatomy
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Spinal Cord: Cross-sectional Anatomy
Gray Matter and its Components
Central to the gray matter is...