Analysis of a Pulse Rate Variability Measurement Using a Smartphone Camera
András Bánhalmi1, János Borbás2, Márta Fidrich1
1Institute of Informatics, University of Szeged, Szeged, 2 Árpád Square 6720, Hungary.
Smartphone photoplethysmography (PPG) accurately measures pulse rate variability (PRV), comparable to electrocardiography (ECG) heart rate variability (HRV). This validates mobile health apps for large-scale autonomic nervous system monitoring.
Area of Science:
- Biomedical Engineering
- Cardiology
- Digital Health
Background:
- Heart rate variability (HRV) reflects autonomic nervous system activity but lacks clinical validation due to limited data.
- Smartphone applications using photoplethysmography (PPG) offer a potential solution for collecting population-level HRV data.
Purpose of the Study:
- To evaluate a smartphone application's capability in measuring pulse rate variability (PRV) using PPG.
- To compare PPG-derived PRV with electrocardiography (ECG)-derived HRV.
- To establish a baseline by comparing differences among ECG channels.
Main Methods:
- Fifty parallel measurements were conducted using an iPhone 6 (240 Hz) and Cardiax PC-ECG devices.
- Correlation, linear regression, and Bland-Altman analysis were employed to assess PRV and HRV index correspondence.
- Intra-ECG channel variability was analyzed for baseline comparison.
Main Results:
- The smartphone PPG demonstrated high accuracy, with deviations comparable to those between ECG channels.
- Mean deviations between PPG-ECG and ECG channels were minimal for RR (0.01–0.06 ms), SDNN (0.78–0.46 ms), RMSSD (1.79–1.21 ms), and pNN50 (2.43%–1.63%).
Conclusions:
- The iPhone application provides reliable PPG-based PRV indices, comparable to ECG-based HRV.
- These findings support the use of smartphone PPG for accessible autonomic nervous system assessment.
- Further research will explore PPG-ECG correspondence across different ECG channels.
Related Concept Videos
Increased pulse rate
Many factors can elevate the risk of developing tachycardia. These include advanced age, a family history of arrhythmias, and an...
Decreased pulse rate
There are specific risk factors that can elevate the likelihood of developing bradycardia. Advanced age is a significant factor, with...
Variability: Analysis
The range is a simple measure of variability, indicating the difference between the highest and...
Measuring Reaction Rates
Special considerations while measuring pulse
Pulse
The pulse serves as a clinical...