Insulin compatibility with polymer materials used in external pump infusion systems
S G Melberg1, S Havelund, J Villumsen
1Novo Research Institute, Bagsvaerd, Denmark.
Summary
Polyvinylchloride (PVC) infusion sets, especially those sterilized with gamma irradiation, degrade insulin quality by reducing preservatives and increasing harmful chemical transformations. Polyethylene infusion sets sterilized with ethylene oxide maintain insulin stability.
Area of Science:
- Pharmaceutical Science
- Materials Science
Background:
- Insulin pump therapy is a common treatment for diabetes.
- The materials and sterilization methods of infusion sets can potentially affect insulin quality.
Purpose of the Study:
- To assess the impact of different external insulin pump infusion set materials and sterilization methods on insulin quality.
- To identify which infusion set components and processes maintain insulin stability.
Main Methods:
- Insulin quality was analyzed using high-performance liquid chromatography (HPLC).
- Two types of infusion sets were tested: PVC/rubber with PVC catheter (gamma irradiation) and polyethylene/polypropylene with polyethylene catheter (ethylene oxide sterilization).
- Insulin solutions were passed through the sets under simulated user conditions.
Main Results:
- The PVC infusion set significantly reduced preservatives to below 30% and increased insulin transformation products (2-3x baseline).
- The polyethylene/polypropylene infusion system showed minimal preservative loss and no significant increase in insulin transformation.
- Gamma irradiation of PVC catheters was detrimental to insulin stability, while ethylene oxide sterilization of polyethylene catheters did not affect it.
- These effects were independent of insulin brand and zinc content.
Conclusions:
- Gamma-irradiated PVC infusion sets compromise insulin quality, impacting therapeutic efficacy.
- Ethylene oxide-sterilized polyethylene infusion sets are suitable for maintaining insulin stability in external insulin pumps.
- Material selection and sterilization methods are critical factors in insulin pump infusion set design.