Use of impedance-based catheter tip-to-tissue contact assessment (electroanatomic coupling index, ECI) in typical
Massimiliano Maines1, Francesco Peruzza2, Alessandro Zorzi3
1Department of Cardiology, Santa Maria del Carmine Hospital, Corso Verona 4, 38068, Rovereto, Trento, Italy. massimiliano.maines@apss.tn.it.
Purpose:
The electrical coupling index (ECI) (Abbott, USA) is a marker of tissue contact and ablation depth developed particularly for atrial fibrillation treatment. We sought to evaluate if these measures can be also a marker of lesion efficacy during cavotricuspid isthmus (CTI) ablation for typical right atrial flutter.
Methods:
We assessed the ECI values in patients undergoing typical right atrial flutter point-by-point ablation guided by the Ensite Velocity Contact™ (St. Jude Medical, now Abbott St. Paul, MN, USA) electroanatomic mapping system. ECI values were collected before, during (at the plateau), and after radiofrequency (RF) delivery. The physician was blinded to ECI and judged ablation efficacy according to standard parameters (impedance drop, local potential reduction, and/or split in two separate potentials). Patients were followed up at 3 and 12 months.
Results:
Fifteen consecutive patients (11 males, mean age 69.2 ± 10.6 years) with a history of typical right atrial flutter were included in this study. A total of 158 RF applications were assessed (mean 10.5 ± 6.6 per patient, range 6-28). The absolute and percentage ECI variations (pre-/post-ablation) were significantly greater when applications were effective (p < 0.001). A 12% drop in the ECI after ablation was identified by the ROC curve as the best cutoff value to discriminate between effective and ineffective ablation (sensitivity 94%, specificity 100%). Acute success was achieved in all patients with no complications and no recurrences during follow-up.
Conclusion:
The ECI appeared a reliable index to guide CTI ablation. A 12% drop of ECI during radiofrequency energy delivery was highly accurate in identifying effective lesion.
Related Concept Videos
Transmission-based Precautions I: Contact, Enteric, and Droplets
Contact Precautions:
Contact precautions are the measures taken to prevent the transmission of infectious agents, especially epidemiologically important microorganisms such as MRSA or influenza, primarily transmitted through direct or indirect contact with an...
Typical Model Studies
Contact Angle
The adhesive force is the molecular force between molecules of different materials, that is, between the molecules of the solid and the liquid. The cohesive...
Contact-dependent Signaling
Gap Junctions
In animal cells, gap junctions are formed...
Impedance Combination
Impedances and Admittance
Impedance is a measure of resistance to sinusoidal current flow in an AC circuit. Unlike their behavior in DC circuits, where inductors appear as short circuits and capacitors as open circuits, the behavior of these components in AC...