Usefulness of Reflection Scanning in Determining Whole-Body Composition in Broadly Built Individuals Using
Andreia V Moço1, Catarina N Matias1, Diana A Santos1
1Exercise and Health Laboratory, CIPER, Faculdade Motricidade Humana, Universidade de Lisboa, Cruz Quebrada, Portugal.
Abstract:
Whole-body composition analysis by dual-energy X-ray absorptiometry (DXA) requires subjects to fit within the width limits of the DXA bed. To overcome this limitation, the aim of this study was to validate a partial scanning technique at which the upper left limb is deliberately left unscanned and measurements are "reflected" from the right-side upper limb. A Hologic Explorer-W densitometer was used in a sample of 189 participants, including athletes and nonathletes, ranging from underweight to obese (body mass index: 17.0-40.1 kg/m2). A whole-body scan was analyzed as the reference procedure to determine bone mineral content (BMC), lean soft tissue (LST), and fat mass (FM), and reanalyzed using a partial reflection scanning (RS) technique. RS estimates of BMC were associated with athletic status and differed significantly from reference estimates (p < 0.05). Also, the RS estimates of LST and FM were different (p < 0.05) from those of the reference whole-body scan, although differences were small (0.17 kg, -0.02 kg, and -0.10% for BMC, LST, and FM, respectively). The alternative procedure explained more than 99% of the reference scan variance with low limits of agreement (BMC: -13.8 to 23.9 g [athletes] and -6.3 to 18.0 g [nonathletes]; LST: -0.11 to 0.45 kg; FM: -0.22 to 0.17 kg). Regardless of body mass index, athletic status, and gender, RS is a useful and simple solution to be used in individuals wider than the DXA scan area. However, individual errors for BMC may be higher in athletes engaged in lateral dominant sports practice.
More Related Videos
08:22Using Deuterium Oxide as a Non-Invasive, Non-Lethal Tool for Assessing Body Composition and Water Consumption in Mammals
Published on: February 20, 2020
06:27Expression of Cementitious Pore Solution and the Analysis of Its Chemical Composition and Resistivity Using X-ray Fluorescence
Published on: September 23, 2018
Related Concept Videos
Composite Bodies
Composite bodies have widespread applications in mechanical engineering, from automobiles to aircraft to rockets. For example, an automobile wheel comprises...
Composition of Body Fluids
Kinetic Energy for a Rigid Body
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
Free Energy Changes for Nonstandard States
What is Energy?