Published on: July 12, 2022
Are Radiographic and Direct Measures of Acetabular Polyethylene Wear Comparable?
Krista K Parran1, Christopher P Bechtel2, Rebecca D Moore1
1Department of Orthopaedics, Case Western Reserve University School of Medicine, Cleveland, OH.
Direct measurements of polyethylene acetabular liner wear correlate well with the computerized Hip Analysis Suite (HAS) method, especially when comparing maximum wear. A single prerevision radiograph is sufficient for assessing in vivo wear.
Area of Science:
- Orthopedic surgery
- Biomaterials science
- Medical imaging analysis
Background:
- Polyethylene acetabular liners in hip implants are prone to wear over time.
- Various methods exist for calculating linear wear rates, necessitating validation.
Purpose of the Study:
- To compare 2-dimensional wear rates derived from direct micrometer measurements against those from the computerized Hip Analysis Suite (HAS) edge-detection method.
- To validate the accuracy of HAS in assessing polyethylene wear in retrieved acetabular liners.
Main Methods:
- Evaluated two groups of retrieved acetabular liners (n=18 and n=55) from Harris-Galante Prosthesis implants with over 10 years in situ.
- Calculated average and maximum direct linear wear using a digital micrometer at six locations.
- Utilized HAS 8.0.4.3 to compute 2-dimensional wear from anteroposterior pelvic radiographs.
Main Results:
- HAS-calculated aggregate wear rates were significantly higher than average direct measurements for both groups (P = .020 and P < .001).
- No significant difference was found when comparing maximum direct micrometer measurements to HAS for either group (P = .351 and P = .451).
- Strong linear correlation was observed between HAS and both average and maximum direct wear measures, with maximum wear showing closer correspondence to HAS.
Conclusions:
- This study is the first to compare and validate 2-dimensional wear rates using direct measurements from retrieved components against a radiographic computer-assisted technique.
- Wear rates determined by direct measurements are consistent with computer-assisted 2-dimensional methods, particularly for maximum wear.
- A single prerevision radiograph is adequate for assessing 2-dimensional in vivo wear of acetabular liners.
Related Concept Videos
Microbial Growth Measurement: Direct Methods
Comparative Excretory Systems
Comparing Copy Number Variations and SNPs
Copy number variations or CNVs are the structural variations that cover more than 1kb of DNA sequence. The single nucleotide polymorphism (SNP), on the other hand, is a single nucleotide change or a point mutation that is found in more than 1%...
Comparing Experimental Results: Student's t-Test
Comparing the Survival Analysis of Two or More Groups
Directing Effect of Substituents: meta-Directing Groups