Evaluation of a Portable Urinary pH Meter and Reagent Strips
Vincent De Coninck1,2, Etienne Xavier Keller1,3, María Rodríguez-Monsalve1
11 GRC n°20, Groupe de Recherche Clinique sur la Lithiase Urinaire, Sorbonne Université , AP-HP, Hôpital Tenon, Paris, France .
Objective:
To evaluate a portable electronic pH meter and to put its accuracy in perspective with reagent strips read by a layperson, a healthcare professional, and an electronic reading device.
Materials And Methods:
Based on a preanalysis on 20 patients, a sample size of 77 urine aliquots from healthy volunteers was necessary to obtain sufficient study power. Measurements of urinary pH were obtained by use of reagent strips, a portable pH meter and a laboratory pH meter (gold standard). Reagents strips were read by a professional experienced in interpreting strips, a layperson, and an electronic strip reader. The mean matched pair difference between measurement methods was analyzed by the paired t-test. The degree of correlation and agreement were evaluated by the Pearson's correlation coefficient and Bland-Altman plots, respectively.
Results:
The mean matched pair difference between the gold standard and all other pH measurement methods was the smallest with the portable electronic pH meter (bias 0.01, 95% confidence interval [CI] -0.07 to 0.08; p = 0.89), followed by strips read by a professional (bias -0.09, 95% CI -0.21 to 0.02; p = 0.10), layperson (bias -0.17, 95% CI -0.31 to -0.04; p = 0.015), and electronic strip reader (bias -0.29, 95% CI -0.41 to -0.16; p < 0.001). The portable electronic pH meter achieved the highest Pearson's correlation coefficient and narrowest 95% limits of agreement, followed by strip interpretation by a professional, electronic strip reader, and layperson. To quantify the ability of pH measurement methods to correctly classify values within a predefined urinary pH target range, we performed classification tests for several stones. The portable electronic pH meter outperformed all other measurement methods for negative predictive values.
Conclusions:
Findings of this study support that the portable electronic pH meter is a reliable pH measuring device. It appears to be more accurate compared to reagent strips readings.
More Related Videos
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
08:46Rapid Diagnosis of Avian Influenza Virus in Wild Birds: Use of a Portable rRT-PCR and Freeze-dried Reagents in the Field
Published on: August 2, 2011
Related Concept Videos
Voltammetry: Stripping Methods
Anodic Stripping Voltammetry (ASV)
ASV is used to determine metals and metalloids at trace levels. It involves two steps: deposition and stripping. First, a negative potential is applied to the...
EDTA: Auxiliary Complexing Reagents
Introduction to Urinary System
The kidneys are bean-shaped organs located in the retroperitoneal space, on either side of the vertebral column, between the T12 and L3 vertebrae. They are partially protected by the rib cage and surrounded by perirenal fat, which provides cushioning. They are responsible for urine formation and play critical roles in regulating blood pressure, electrolyte levels, and hormone production. The ureters...
Urinary Bladder
In males, the bladder is situated in front of the rectum, while in females, it is positioned anterior to the vagina and uterus. The bladder floor contains an inverted triangular area called the trigone, defined by the two ureteric...
Disorders of the Urinary System
Urinary tract infections (UTIs) are one of the most common urinary system disorders. They are caused by bacteria that enter the urethra and can spread to the bladder resulting in cystitis. Pyelonephritis is the result of a UTI that has ascended to the level of the...
Acid Halides to Ketones: Gilman Reagent
As shown below, the mechanism proceeds in two steps. First, one of the alkyl groups of the reagent acts as a nucleophile and attacks the acyl carbon of the acid chloride to form a tetrahedral intermediate. This is followed by the reformation of the carbon–oxygen...