In Vitro Performance Testing of Legacy and ENFit Gastrostomy Tube Devices Under Gravity Flow Conditions
Suvajyoti Guha1, Nivetita Ravi1,2, Joshua S Silverstein3
1Office of Science and Engineering Laboratories, Center for Devices and Radiological Health, U.S. Food and Drug Administration, Silver Spring, Maryland, USA.
Background:
Changes in connector design for the gastrostomy tube were implemented to reduce the risk of misconnections. This study aimed to determine whether there are differences in gravity flow rates between legacy devices and the ENFit devices intended to replace them.
Materials And Methods:
We compared 5 legacy gastrostomy tube brands with 3 corresponding ENFit brands, sized between 14 French (Fr) and 24 Fr. Seven commercial diets were used. One comparison involved low-profile devices.
Results:
Whether an ENFit device manifested a lower flow rate than a legacy device was not a strong function of diet. One 14-Fr ENFit device, because of its reduced distal inner tube diameter, produced an average feeding time of 56 (±13) minutes from a 20-minute baseline. For other 14-Fr ENFit devices, the increase was much less pronounced (25 ± 4 minutes). At larger sizes, both decreases and increases in feeding time were observed, depending on device type; on average, the 20-minute feeding time increased to 25 (±7) minutes. For low-profile devices, across all sizes, an increase in 20-minute feeding time occurred, but the difference was small (23 ± 2).
Conclusion:
Statistically lower flow rates were observed for 70% of ENFit devices relative to their legacy counterparts. We estimate that 30% of the differences may be noticeable. In the scenarios studied, lower flow rates (relative to other devices at the same Fr number) arise from energy losses in straight tubing. This difference can be reduced by increasing the tube inner diameters in distal end of ENFit tubes.
More Related Videos
Related Concept Videos
Steady, Laminar Flow in Circular Tubes
Responses to Gravity and Touch
Center of Gravity
Center of Gravity
To determine its location, the principle of moments can be utilized by dividing the object into...
Enteral Nutrition II: Nasointestinal and Gastrostomy Feeding
Nasointestinal Feeding
Nasointestinal feeding involves placing a tube...
Work Done by Gravity
Like other forces, gravity does work on an object if it displaces it toward the Earth's center. In this case, the work done by gravity is said to be positive. If an external force acts on the object against the pull of gravity and manages to lift it away from the Earth's center, work is done against gravity. In this case, the net work done is said to be...