Dosimetry for Cell Irradiation using Orthovoltage 40-300 kV X-Ray Facilities
Published on: February 20, 2021
Importance of dosimetry protocol for cell irradiation on a low X-rays facility and consequences for the biological
Morgane Dos Santos1, Vincent Paget2, Mariam Ben Kacem2
1a Department of RAdiobiology and Regenerative MEDicine (SERAMED), Laboratory of Radiobiology of Accidental Exposures (LRAcc) , Institute for Radiological Protection and Nuclear Safety (IRSN) , Fontenay-aux-Roses , France.
Purpose:
The main objective of radiobiology is to establish links between doses and radiation-induced biological effects. In this context, well-defined dosimetry protocols are crucial to the determination of experimental protocols. This work proposes a new dosimetry protocol for cell irradiation in a SARRP and shows the importance of the modification of some parameters defined in dosimetry protocol for physical dose and biological outcomes.
Materials And Methods:
Once all parameters of the configuration were defined, dosimetry measurements with ionization chambers and EBT3 films were performed to evaluate the dose rate and the attenuation due to the cell culture medium. To evaluate the influence of changes in cell culture volume and/or additional filtration, 6-well plates containing EBT3 films with water were used to determine the impact on the physical dose at 80 kV. Then, experiments with the same irradiation conditions were performed by replacing EBT3 films by HUVECs. The biological response was assessed using clonogenic assay.
Results:
Using a 0.15 mm copper filter lead to a variation of +1% using medium thickness of 0.104 cm to -8% using a medium thickness of 0.936 cm on the physical dose compare to the reference condition (0.313 cm). For the 1 mm aluminum filter, a variation of +8 to -40% for the same medium thickness conditions has been observed. Cells irradiated in the same conditions showed significant differences in survival fraction, corroborating the effects of dosimetric changes on physical dose.
Conclusions:
This work shows the importance of dosimetry in radiobiology studies and the need of an accurate description of the dosimetry protocol used for irradiation.
Related Concept Videos
X-ray Diffraction of Biological Samples
According to Bragg's law, when X-rays strike the sample positioned on a stage, the rays are scattered by the electron clouds around the sample atoms. The X-ray diffraction or scattering is caused by constructive interference of the X-ray waves that reflect off the internal...
Biological Clocks and Seasonal Responses
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
Biological Effects of Radiation
Timing and Consequences on Behavior
Humans, however, can respond to delayed reinforcers. We often make decisions between immediate small rewards and delayed larger rewards. This ability to delay gratification is a significant...
What is Conservation Biology?