Investigation of EBT3 radiochromic film's response to humidity
Elsa Y León-Marroquín1, José M Lárraga-Gutiérrez2, J Alfredo Herrera-González2
1Laboratorio de Fotomedicina, Biofotónica y Espectroscopia Láser de Pulsos Ultracortos, Facultad de Medicina, Universidad Autónoma del Estado de México, Toluca, México.
Purpose:
The aim of this work is to investigate the effects of immersing EBT3 radiochromic film in water and to evaluate its contribution to the total uncertainty in dose determination.
Materials And Methods:
We used 3 cm × 3 cm EBT3 radiochromic films irradiated in the range of 0-70 Gy to study the impact of water immersion on the change in net optical density. These films were placed in a water container for a period of 24 h. The net optical density was measured before (0 h) and after of the immersion in water (1, 3, 6, 12, 18, and 24 h). The absorbance spectrum of the EBT3 radiochromic film was measured at 0 h and 24 h after immersion in water. The uncertainty in dose determination due to the effects of keeping the EBT3 radiochromic film submerged in water at 0, 1, and 24 h were recorded in the red, green, and blue channels.
Results:
We observed an increase in the net optical density as an effect on the film due to its immersion in water. The penetration of the water at the edges of the radiochromic film was observed to be a function of time during which the film remained in the water. On the other hand, the penetration of water at the edges of the film was found to be independent of irradiation dose.
Conclusions:
EBT3 radiochromic film is found more resistant to water penetration through the edges than its predecessors. However, there is evidence that suggest that liquid water damage the Nylon cover layer of the film by changing its optical properties. Therefore, it is recommended to build a new calibration curve for radiochromic films for a specific situation involving dose measurements in liquid water.