The Potential Role of Grid-Like Software in Bedside Chest Radiography in Improving Image Quality and Dose Reduction:
Su Yeon Ahn1,2, Kum Ju Chae1,3, Jin Mo Goo1,4
1Department of Radiology, Seoul National University College of Medicine, and Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul 03080, Korea.
Objective:
To compare the observer preference of image quality and radiation dose between non-grid, grid-like, and grid images.
Materials And Methods:
Each of the 38 patients underwent bedside chest radiography with and without a grid. A grid-like image was generated from a non-grid image using SimGrid software (Samsung Electronics Co. Ltd.) employing deep-learning-based scatter correction technology. Two readers recorded the preference for 10 anatomic landmarks and the overall appearance on a five-point scale for a pair of non-grid and grid-like images, and a pair of grid-like and grid images, respectively, which were randomly presented. The dose area product (DAP) was also recorded. Wilcoxon's rank sum test was used to assess the significance of preference.
Results:
Both readers preferred grid-like images to non-grid images significantly (p < 0.001); with a significant difference in terms of the preference for grid images to grid-like images (p = 0.317, 0.034, respectively). In terms of anatomic landmarks, both readers preferred grid-like images to non-grid images (p < 0.05). No significant differences existed between grid-like and grid images except for the preference for grid images in proximal airways by two readers, and in retrocardiac lung and thoracic spine by one reader. The median DAP were 1.48 (range, 1.37-2.17) dGy*cm2 in grid images and 1.22 (range, 1.11-1.78) dGy*cm2 in grid-like images with a significant difference (p < 0.001).
Conclusion:
The SimGrid software significantly improved the image quality of non-grid images to a level comparable to that of grid images with a relatively lower level of radiation exposure.
More Related Videos
08:59Cryo-Electron Microscopic Grid Preparation for Time-Resolved Studies using a Novel Robotic System, Spotiton
Published on: February 25, 2021
08:22Stereotaxic Microinjection of Viral Vectors Expressing Cre Recombinase to Study the Role of Target Genes in Cocaine Conditioned Place Preference
Published on: July 30, 2013
Related Concept Videos
Observational Studies
There are three types of observational studies – Prospective, retrospective, and cross-sectional.
Prospective Study
Prospective studies, also known as longitudinal or cohort studies, are carried out by collecting future data from groups sharing similar characteristics. One...
Oxidation-Reduction Reactions
Flail Chest-I
Flail chest is a severe and potentially life-threatening condition characterized by the fracture of three or more adjacent ribs in multiple places. It is most commonly caused by direct impacts and trauma, such as motor vehicle accidents or injuries from a steering wheel impact. It can also occur due to falls in elderly individuals with osteoporosis, or assaults involving sharp objects.
Pathophysiology
The pathophysiology of flail chest is complex, involving fractures of...
Flail Chest-II
Assessment:
1. Clinical Evaluation:
History:
Chest Physiotherapy
Purpose
CPT is primarily used for patients with excessive bronchial secretions who have difficulty clearing...
Bioavailability Study Design: Single Versus Multiple Dose Studies