Impact of the Lok-bar for High-precision Radiotherapy with Tomotherapy
Makoto Hirata1, Hajime Monzen2, Mikoto Tamura1,3
1Department of Medical Physics, Graduate School of Medical Science, Kindai University, Osaka, Japan.
Background/Aim:
Patient immobilization systems are used to establish a reproducible patient position relative to the couch. In this study, the impact of conventional lok-bars for CT-simulation (CIVCO-bar) and treatment (iBEAM-bar) were compared with a novel lok-bar (mHM-bar) in tomotherapy.
Materials And Methods:
Verification was obtained as follows: i. artifacts in CT images; ii. dose attenuation rate of lok-bar, compared to without lok-bar; and iii. dose differences between the calculated and measured absorbed doses.
Results:
With the CIVCO-bar, there were obvious metal artifacts, while there were nearly no artifacts with the mHM-bar. The mean dose attenuation rates with the mHM-bar and iBEAM-bar were 1.31% and 2.28%, and the mean dose difference was 1.55% and 1.66% for mHM-bar and iBEAM-bar.
Conclusion:
Using the mHM-bar reduced artifacts on the CT image and improved dose attenuation are obtained. The lok-bar needs to be inserted as a structure set in treatment planning with tomotherapy.
Related Concept Videos
Bar Graph
Multiple Bar Graph
Each bar or column in the multiple bar graph represents a data value. These graphs are used primarily in interrelating two or more sets of data. The categories of different kinds of data are listed along the horizontal or x-axis, whereas...
Uncertainty in Measurement: Accuracy and Precision
Impact of Groups on Groups
Impact
When particles with different initial velocities collide, they induce deformation by applying equal and opposite impulses. At the point of maximum deformation, the particles move together with...
Accuracy and Precision