CT metal artifacts in patients with total hip replacements: for artifact reduction monoenergetic reconstructions and
Kai Roman Laukamp1, Simon Lennartz1, Victor-Frederic Neuhaus1
1Department of Radiology, University of Cologne, Kerpener Straße 62, 50937, Cologne, Germany.
Objectives:
This study compares metal artifact (MA) reduction in imaging of total hip replacements (THR) using virtual monoenergetic images (VMI), for MA-reduction-specialized reconstructions (MAR) and conventional CT images (CI) from detector-based dual-energy computed tomography (SDCT).
Methods:
Twenty-seven SDCT-datasets of patients carrying THR were included. CI, MAR and VMI with different energy-levels (60-200 keV) were reconstructed from the same scans. MA width was measured. Attenuation (HU), noise (SD) and contrast-to-noise ratio (CNR) were determined in: extinction artifact, adjacent bone, muscle and bladder. Two radiologists assessed MA-reduction and image quality visually.
Results:
In comparison to CI, VMI (200 keV) and MAR showed a strong artifact reduction (MA width: CI 29.9±6.8 mm, VMI 17.6±13.6 mm, p<0.001; MAR 16.5±14.9 mm, p<0.001; MA density: CI -412.1±204.5 HU, VMI -279.7±283.7 HU; p<0.01; MAR -116.74±105.6 HU, p<0.001). In strong artifacts reduction was superior by MAR. In moderate artifacts VMI was more effective. MAR showed best noise reduction and CNR in bladder and muscle (p<0.05), whereas VMI were superior for depiction of bone (p<0.05). Visual assessment confirmed that VMI and MAR improve artifact reduction and image quality (p<0.001).
Conclusions:
MAR and VMI (200 keV) yielded significant MA reduction. Each showed distinct advantages both regarding effectiveness of artifact reduction, MAR regarding assessment of soft tissue and VMI regarding assessment of bone.
Key Points:
• Spectral-detector computed tomography improves assessment of total hip replacements and surrounding tissue. • Virtual monoenergetic images and MAR reduce metal artifacts and enhance image quality. • Evaluation of bone, muscle and pelvic organs can be improved by SDCT.
More Related Videos
07:20A Novel Approach to Overcome Movement Artifact When Using a Laser Speckle Contrast Imaging System for Alternating Speeds of Blood Microcirculation
Published on: August 30, 2017
10:23Equipment Setup and Artifact Removal for Simultaneous Electroencephalogram and Functional Magnetic Resonance Imaging for Clinical Review in Epilepsy
Published on: June 23, 2023
Related Concept Videos
Oxidation-Reduction Reactions
Metal-Ligand Bonds
In these complexes, transition metals form coordinate covalent bonds, a kind of Lewis acid-base interaction in which both of the electrons in the bond are contributed by a donor (Lewis base) to an electron acceptor (Lewis acid). The Lewis acid in...
Metallic Solids
All metallic solids exhibit high thermal and electrical conductivity, metallic luster, and malleability....
Bonding in Metals
Alkali Metals
Table 1: Properties of the alkali metals
Nursing Process for Patient and Caregiver Teaching II: Planning and Implementation