Cerebral Protection During Transcatheter Aortic Valve Implantation: An Updated Systematic Review and Meta-Analysis
Luca Testa1, Azeem Latib2, Matteo Casenghi3
1Department of Cardiology, IRCCS Pol. S. Donato, S. Donato Milanese Milan, Italy luctes@gmail.com.
Insights
Embolic protection devices (EPD) may reduce stroke during transcatheter aortic valve implantation but do not affect mortality or lesion count. EPD use is linked to smaller lesion volumes, suggesting improved safety.
Area of Science:
- Cardiology
- Interventional Cardiology
- Neurology
Background:
- Transcatheter aortic valve implantation (TAVI) carries a risk of cerebral embolism.
- Embolic protection devices (EPD) are designed to mitigate this risk.
- Existing evidence on EPD efficacy in TAVI is limited and inconclusive.
Purpose of the Study:
- To evaluate the safety and efficacy of EPD during TAVI.
- To assess the impact of EPD on mortality, stroke, and cerebral lesion burden.
Main Methods:
- A meta-analysis of eight studies involving 1285 patients undergoing TAVI.
- Searched major medical databases up to December 2017.
- Compared outcomes between patients with and without EPD use.
Main Results:
- EPD delivery success rate was 94.5%.
- No significant difference in 30-day mortality (OR 0.43 [0.18-1.05], P=0.3).
- Associated with a lower 30-day stroke rate (OR 0.55 [0.31-0.98], P=0.04).
- No difference in the number of new lesions (SMD -0.19 [-0.71 to 0.34], P=0.49).
- Associated with significantly smaller ischemic volume per lesion (SMD -0.52 [-0.85 to -0.20], P=0.002) and total lesion volume (SMD -0.23 [-0.42 to -0.03], P=0.02).
Conclusions:
- EPD use in TAVI is not associated with reduced mortality or new ischemic lesions.
- EPD may be associated with a lower 30-day stroke rate, though this finding requires further validation.
- EPD use is linked to a reduction in the volume of ischemic cerebral lesions.
Background:
The use of embolic protection devices (EPD) may theoretically reduce the occurrence of cerebral embolic lesions during transcatheter aortic valve implantation. Available evidence from single studies is inconclusive. The aim of the present meta-analysis was to assess the safety and efficacy profile of current EPD.
Methods And Results:
Major medical databases were searched up to December 2017 for studies that evaluated patients undergoing transcatheter aortic valve implantation with or without EPD. End points of interest were 30-day mortality, 30-day stroke, the total number of new lesions, the ischemic volume per lesion, and the total volume of lesions. Eight studies involving 1285 patients were included. The EPD delivery success rate was reported in all studies and was achieved in 94.5% of patients. The use of EPD was not associated with significant differences in terms of 30-day mortality (odds ratio 0.43 [0.18-1.05], P=0.3) but it was associated with a lower rate of 30-day stroke (odds ratio 0.55 [0.31-0.98], P=0.04). No differences were detected with respect to the number of new lesions (standardized mean difference -0.19 [-0.71 to 0.34], P=0.49). The use of EPD was associated with a significantly smaller ischemic volume per lesion (standardized mean difference, -0.52 [-0.85 to -0.20], P=0.002) and smaller total volume of lesions (standardized mean difference, -0.23 [-0.42 to -0.03], P=0.02).
Conclusions:
The use of EPD is not associated with a reduced rate of mortality and new ischemic cerebral lesions. The use of EPD during transcatheter aortic valve implantation seems to be associated with a lower 30-day stroke rate, although this result is driven by a single nonrandomized study. The use of EPD is associated with a smaller volume of ischemic lesions, and smaller total volume of ischemic lesions.
Related Concept Videos
Review and Preview
Percentiles are a type of fractile that partition data into...
Review and Preview
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Random and Systematic Errors
Directing Effect of Substituents: meta-Directing Groups
Systematic Sampling Method
Systematic sampling is one of the simplest methods...