Suction forces generated by passive bile bag drainage on a model of post-subdural hematoma evacuation
Steven O Tenny1, William E Thorell2
1Division of Neurosurgery, Department of Surgery, University of Nebraska Medical Center, 982035 Nebraska Medical Center, Omaha, NE, 68198, USA. steven.tenny@unmc.edu.
Background:
Passive drainage systems are commonly used after subdural hematoma evacuation but there is a dearth of published data regarding the suction forces created. We set out to quantify the suction forces generated by a passive drainage system.
Method:
We created a model of passive drainage after subdural hematoma evacuation. We measured the maximum suction force generated with a bile bag drain for both empty drain tubing and fluid-filled drain tube causing a siphoning effect. We took measurements at varying heights of the bile bag to analyze if bile bag height changed suction forces generated.
Results:
An empty bile bag with no fluid in the drainage tube connected to a rigid, fluid-filled model creates minimal suction force of 0.9 mmHg (95% CI 0.64-1.16 mmHg). When fluid fills the drain tubing, a siphoning effect is created and can generate suction forces ranging from 18.7 to 30.6 mmHg depending on the relative position of the bile bag and filled amount of the bile bag. The suction forces generated are statistically different if the bile bag is 50 cm below, level with or 50 cm above the experimental model.
Conclusion:
Passive bile bag drainage does not generate significant suction on a fluid-filled rigid model if the drain tubing is empty. If fluid fills the drain tubing then siphoning occurs and can increase the suction force of a passive bile bag drainage system to levels comparable to partially filled Jackson-Pratt bulb drainage.
Related Concept Videos
Bile
Bile is released when dietary fats enter...
Tracheostomy Suctioning II: Procedure
Suctioning the Nasopharyngeal Airway
Equipment Required
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Passive Filters
Low-Pass Filters
Low-pass filters are designed to transmit signals with frequencies lower than the cutoff frequency, ωc, and attenuate those above it. The cutoff...
Active versus Passive Immunity
Active Immunity
Active immunity refers to the resistance one develops...