Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
A novel-integrated quality assurance phantom for radiographic and nonradiographic radiotherapy localization and
Amy S Yu1, Tyler L Fowler1, Piotr Dubrowski1
1Department of Radiation Oncology, Stanford University School of Medicine, Stanford, CA, USA.
Purpose:
Various localization and positioning systems utilizing radiographic or nonradiographic methods have been developed to improve the accuracy of radiation treatment. Each quality assurance (QA) procedure requires its own phantom and is independent from each other, so the deviation between each system is unavailable. The purpose of this work is to develop and evaluate a single-integrated QA phantom for different localization and positioning systems.
Methods:
The integrated phantom was designed in three-dimensional (3D) CAD software and 3D printed. The phantom was designed with laser alignment marks, a raised letter "S" on the anterior surface for optical surface monitoring system registration, a core for radiofrequency (RF) tracking system alignment, eight internal fiducials for image alignment, and an isocentric bearing for Winston-Lutz test. Tilt legs and rotational stage were designed for rotational verification of optical surface mapping system and RF tracking system, respectively. The phantom was scanned using a CT scanner and a QA plan was created. This prototype phantom was evaluated against established QA techniques.
Results:
The QA result between the proposed procedure and established QA technique are 1.12 ± 0.31 and 1.14 ± 0.31 mm, respectively, for RF tracking system and 0.18 ± 0.06 and 0.18 ± 0.05 mm for Winston-Lutz test. There is no significant difference for the QA results between the established QA and proposed procedure (P > 0.05, t test). The accuracy of rotational verification for surface mapping system and RF tracking system are less than 0.5 and 1° compared the predefined value. The isocenter deviation of each location system is around l mm.
Conclusion:
We have designed and evaluated a novel-integrated phantom for radiographic and nonradiographic localization and positioning systems for radiotherapy. With this phantom, we will reduce the variation in measurements and simplify the QA procedures.
Related Concept Videos
Quality Assurance
Velocity and Position by Integral Method
Consider an example to calculate the velocity and position from the acceleration function. A motorboat is traveling at a constant velocity of 5.0 m/s when it starts to decelerate to arrive at the dock. Its acceleration is...
Root Loci for Positive-Feedback Systems
The construction rules for the root locus in positive feedback systems are similar to those in...
Position-effect Variegation
Quality Control
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...
Quality of Water