Related Experiment Video
Updated: Feb 11, 2026

Cell-based Assay Protocol for the Prognostic Prediction of Idiopathic Scoliosis Using Cellular Dielectric Spectroscopy
Published on: October 16, 2013
Reliability of Three-Dimensional Spinal Modeling of Patients With Idiopathic Scoliosis Using EOS System
Aria Bagheri1, Xue-Cheng Liu2, Channing Tassone2
1Medical College of Wisconsin, 8701 Watertown Plank Road, Milwaukee, WI 53226, USA.
Study Design:
Three-dimensional (3D) spinal models of children with idiopathic scoliosis (IS) were created using the EOS imaging system (EOS) and sterEOS software.
Objective:
To determine the inter- or intraobserver reproducibility of the 3D spinal models in children with IS of different apex locations.
Summary Of Background Data:
3D spinal model measurements include the Cobb angle, kyphosis, lordosis, and axial vertebral rotation (AVR). Variation of these measurements between two investigators and two different trials by the same investigator were analyzed by inter- and intraclass correlation coefficients (ICCs).
Methods:
Biplanar radiographic images of 15 patients (age: 6-15 years) with IS were uploaded into the sterEOS software. Spinal and pelvic markers were manually identified to construct a 3D spinal model and measure spinal parameters. Two trained examiners independently performed modeling and performed modeling in spaced out trials. The ICC between inter- and intraobservers were calculated.
Results:
ICCs between inter- and intraobservers were significant for all parameters (p < .05). Both the inter- and intraobservers showed excellent agreement for the Cobb angles in the thoracic segment, kyphosis and lordosis. Substantial interobserver agreement and excellent intraobserver agreement were determined for the Cobb angle in the thoracolumbar or lumbar (TL/L) segment, with less than 6° difference between two raters and less than 2° difference between two trials. Substantial interobserver agreement for the AVR in the TL/L region and substantial interobserver agreement for the AVR in the thoracic region were found, with less than 4° difference between raters. One rater had substantial intraobserver agreement for the AVR in the TL/L region whereas another rater reported moderate to substantial intraobserver agreement in both the thoracic and TL/L regions, with less than 3° difference between trials.
Conclusion:
The EOS system shows reliable and repeatable results in 3D spinal modeling of children with IS.
Level Of Evidence:
Level III.
Related Concept Videos
Reliability and Validity
Distribution Reliability and Automation
Dimensional Analysis
Conversion Factors and Dimensional Analysis
The unit...
Dimensional Analysis
In fluid mechanics, dimensional...
Spinal Cord
The Spinal Cord