Related Experiment Video
Updated: Feb 10, 2026

Author Spotlight: Advancements in Impedance Monitoring for Cochlear Implant Surgery
Published on: August 4, 2023
Safe Intraoperative Neurophysiologic Monitoring During Posterior Spinal Fusion in a Patient With Cochlear Implants
Godwin Abiola1, Bryan Kevin Ward2, Stephen Bowditch2
1Johns Hopkins School of Medicine.
Objective:
Cochlear implants are generally considered a contraindication for any procedure requiring electrical stimulation near the implant. We present a case of a patient undergoing intraoperative transcranial electrical motor-evoked potential monitoring with a cochlear implant without adverse outcomes.
Patient:
A 12-year-old girl with a history of VACTERL presented with worsening congenital kyphosis and bilateral severe-to-profound hearing loss. Since age 7 the patient used a cochlear implant in the right ear and hearing aid in the left ear. Physical examination and magnetic resonance imaging in 2016 revealed a left-sided 66-degree thoracolumbar kyphosis at T11 making the patient a candidate for surgical correction.
Interventions:
She underwent a posterior spinal fusion surgery, performed with intraoperative transcranial electrical motor-evoked potential monitoring. Steps were taken to mitigate electrical stimulation of the patient's cochlear implant.
Main Outcome Measures:
Postoperative impedance of individual channels, audiometry, and neural response testing were compared with preoperative measurements.
Results:
Significant (>10%) impedance changes were observed postoperatively in channels 1, 2, 4, and 6; however, the net variation across all the channels was low (3%). The patient reported no hearing changes, and no significant changes in hearing threshold were seen in postoperative audiometric testing or neural response testing.
Conclusion:
We present a case of successful posterior spinal fusion with intraoperative neurophysiological monitoring via transcranial electrical stimulation, in a patient with a cochlear implant. With proper precautions, motor-evoked potential monitoring can be safely performed in a patient with a cochlear implant.
Related Concept Videos
Nuclear Fusion
A helium nucleus has a mass that is 0.7% less than that of four hydrogen nuclei; this lost mass is converted into energy during the fusion. This reaction produces about...
Guidelines and Strategies for Safe Computer Charting
Maintain Confidentiality and Security:
Tagging and Fusion Proteins
SNAREs and Membrane Fusion
SNAREs exist in pairs that symmetrically interact and catalyze the fusion of the lipid bilayers in vesicle and target organelle. v-SNARE in the vesicle membrane are single polypeptide chains that bind to a complementary t-SNARE, composed of 2...
Cranial Bones: Superior and Posterior View
The frontal bone is the single bone that forms the forehead. At its anterior midline, between the eyebrows, there is a slight depression called the glabella. The frontal bone also forms the supraorbital margin of the orbit. Near the middle of this margin is the supraorbital foramen, the opening that provides passage for a sensory nerve to the forehead. The frontal bone is thickened just above each supraorbital margin,...
Spinal Cord