Accuracy of Transcutaneous CO2 Values Compared With Arterial and Capillary Blood Gases
Laura L Lambert1, Melissa B Baldwin1, Cruz Velasco Gonzalez2
1Respiratory Care Services, Arkansas Children's Hospital, Little Rock, Arkansas.
Background:
Transcutaneous monitors are utilized to monitor a patient's respiratory status. Some patients have similar values when comparing transcutaneous carbon dioxide (PtcCO ) values with blood gas analysis, whereas others show extreme variability. A retrospective review of data was performed to determine how accurately PtcCO correlated with CO2 values obtained by arterial blood gas (ABG) or capillary blood gas.
Methods:
To determine whether PtcCO values correlated with ABG or capillary blood gas values, subjects' records were retrospectively reviewed. Data collected included the PtcCO value at the time of blood gas procurement and the ABG or capillary blood gas PCO value. Agreement of pairs of methods (ABG vs PtcCO and capillary blood gas vs PtcCO ) was assessed with the Bland-Altman approach with limits of agreement estimated with a mixed model to account for serial measurements per subject.
Results:
A total of 912 pairs of ABG/PtcCO values on 54 subjects and 307 pairs of capillary blood gas/PtcCO values on 34 subjects were analyzed. The PCO range for ABG was 24-106 mm Hg, and PtcCO values were 27-133 mm Hg. The PCO range for capillary blood gas was 29-108 mm Hg, and PtcCO values were 30-103 mm Hg. For ABG/PtcCO comparisons, the Pearson correlation coefficient was 0.82, 95% CI was 0.80-0.84, and P was <.001. For capillary blood gas/PtcCO comparisons, the Pearson correlation coefficient was 0.77, 95% CI was 0.72-0.81, and P was <.001. For ABG/PtcCO , the estimated difference ± SD was -6.79 ± 7.62 mm Hg, and limits of agreement were -22.03 to 8.45. For capillary blood gas/PtcCO , the estimated difference ± SD was -1.61 ± 7.64 mm Hg, and limits of agreement were -16.88 to 13.66. The repeatability coefficient was about 30 mm Hg.
Conclusions:
Based on these data, capillary blood gas comparisons showed less variation and a slightly lower correlation with PtcCO than did ABG comparisons. After accounting for serial measurements per patient, due to the wide limits of agreement and poor repeatability, the utility of relying on PtcCO readings for this purpose is questionable.
Related Concept Videos
Noble Gases
The elements in group 18 are noble gases (helium, neon, argon, krypton, xenon, and radon). They earned the name “noble” because they were assumed to be nonreactive since they have filled valence shells. In 1962, Dr. Neil Bartlett at the University of British Columbia proved this assumption to be false.
Uncertainty in Measurement: Accuracy and Precision
Molecular Comparison of Gases, Liquids, and Solids
Biodiversity and Human Values
Professional Values
The values that are the foundation of the nursing profession are altruism, autonomy, human dignity, and social justice.
First, altruism refers to the concern for the welfare and well-being of others without personal...
Improving Translational Accuracy