HPLC-based Measurement of Glycated Hemoglobin using Dried Blood Spots Collected under Adverse Field Conditions
Duncan Thomas1, Teresa Seeman2, Alan Potter3
1a Economics Department and Duke Global Health Institute , Duke University , Durham , NC , USA.
Abstract:
Glycated hemoglobin (HbA1c) measured using high-performance liquid chromatography (HPLC) assays with venous blood and dried blood spots (DBS) are compared for 143 paired samples collected in Aceh, Indonesia. Relative to gold-standard venous-blood values, DBS-based values reported by the HPLC are systematically upward biased for HbA1c<8% and the fraction diabetic (HbA1c ≥ 6.5%) is overstated almost five-fold. Inspection of chromatograms from DBS assays indicates the % glycosylated calculated by the HPLC excludes part of the hemoglobin A which is misidentified as a hemoglobin variant. Taking this into account, unbiased DBS-based values are computed using data from the machine-generated chromatograms. When the DBS are collected in a clinic-like setting, under controlled humidity/temperature conditions, the recalculated values are almost identical to venous-based values. When DBS are collected under field conditions, the recalculated values are unbiased, but only about half the HbA1c values are measured reliably, calling into question the validity of the other half. The results suggest that collection conditions, particularly humidity, affect the quality of the DBS-based measures. Cross-validating DBS-based HbA1c values with venous samples collected under exactly the same environmental conditions is a prudent investment in population-based studies.
Related Concept Videos
Hemoglobin
When all four heme groups are bound to oxygen, the resulting molecule is called oxyhemoglobin. As a result, arterial blood...
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Local Anesthetics: Adverse Effects
Once absorbed into the systemic circulation, local anesthetics can affect the organs that depend on the functioning of sodium...
Dry Friction
To illustrate this concept, imagine a wooden crate resting on a rough, non-uniform horizontal surface. When an external force is applied to...
Drying Shrinkage
A portion of this drying shrinkage can be reversed; if the concrete is...