Related Experiment Video
Updated: Feb 10, 2026

Drug-Induced Sleep Endoscopy DISE with Target Controlled Infusion TCI and Bispectral Analysis in Obstructive Sleep Apnea
Published on: December 6, 2016
Validation of a photoplethysmography device for detection of obstructive sleep apnea in the perioperative setting
Philipp Faßbender1, Ali Haddad2, Silja Bürgener2
1Klinik für Anästhesiologie und Intensivmedizin, Universität Duisburg-Essen and Universitätsklinikum Essen, Hufelandstr 55, 45147, Essen, Germany. philipp.fassbender@rub.de.
Abstract:
Obstructive sleep apnea (OSA) is a risk factor for perioperative complications, but many OSA patients present undiagnosed. While polysomnography (PSG) is the "gold standard" for diagnosis, its application is technology-intense, time-consuming, expensive, and requires specialists, often delaying surgery. Thus, miniaturized devices were developed for OSA screening aimed at ruling out major OSA while measuring a lesser number of biological signals. We evaluated the accuracy of a photoplethysmography (PPG)-based device for OSA detection. 48 patients with established or strongly suspected (STOP-Questionnaire) OSA scheduled for surgery underwent in their preoperative nights parallel recordings by PPG and a classic polygraphy (PG) devices (SomnoLab2®). We compared the diagnostic accuracy of the PPG in diagnosing mild [Apnea-/Hypopnea-Index (AHI) 5-14 events/h] and moderate-to-severe OSA (AHI > 15). PPG and PG-derived AHI correlated significantly (r = 0.85, p < 0.0001) and high area under curve (AUC) in receiver operator characteristics (ROC) values were seen for both AHI thresholds (0.93 and 0.95, respectively). For an AHI > 5, sensitivity was 100%, specificity 44%, positive predictive value (PPV) 62%, negative predictive value (NPV) 100%, likelihood ratio (LHR) 1.79, and Cohen κ was 0.43. For an AHI > 15, sensitivity was 92%, specificity 77%, PPV 60%, NPV 96%, LHR 4.04, and Cohen κ was 0.59. In a typical perioperative cohort of confirmed and suspected OSA patients, PPG reliably detected OSA patients while showing some false-positive results. Such devices are helpful for preoperative OSA screening.
Related Concept Videos
Sleep Apnea
The condition is more prevalent among...
Reliability and Validity
Insufficient Sleep and Sleep Deprivation
Sleep deprivation is a more severe form of sleep loss...
Stages of Sleep
Before sleep begins, in wakefulness, the brain exhibits primarily beta waves, which are high in frequency and low in amplitude, indicating alertness...
Understanding Sleep
The circadian rhythm, a nearly 24-hour cycle, is deeply influenced by environmental light cues. Light exposure directly affects the hypothalamus, which in turn regulates...
Data Validation
Key parameters for method validation include: