Comparative safety and efficacy of a dedicated postpartum IUD inserter versus forceps for immediate postpartum IUD
Paul D Blumenthal1, Klaira Lerma1, Renita Bhamrah2
1Stanford Program for International Reproductive Education and Services (SPIRES), Division of Family Planning Services and Research, Department of Obstetrics and Gynecology, School of Medicine, Stanford University, 300 Pasteur Drive, HG 332, Stanford, CA, 94305, USA; Population Services International, 1120 19th Street, NW Suite 600, Washington, DC 20036, USA.
Objective:
To compare postpartum intrauterine device (PPIUD) insertion using a newly developed dedicated PPIUD inserter (inserter) to insertion with modified Kelly placental forceps (forceps).
Study Design:
We randomized 500 women at five Indian academic centers to PPIUD insertion of a Copper T380A with either the inserter or forceps. The primary outcomes were fundal placement (assessed by postinsertion ultrasound), ease of insertion on a 5-point Likert scale and safety (infection/perforation). Secondary outcomes included IUD status at 6-8 weeks postpartum, assessed by string check, ultrasound or x-ray.
Results:
Between September 2015 and July 2016, 241 women in the inserter group and 239 women in the forceps group completed the study, with 20 women excluded due to consent withdrawal (n=4), lost to follow-up (n=10) and protocol deviation (n=6). Median distance from the fundus was similar (p=.36): 5.2 mm (0-130) and 5.8 mm (0-120) in the inserter and forceps groups, respectively. Most providers reported insertion to be easy or very easy with inserter (n=195, 81%) and forceps (n=206, 86%; p=.51). There were no perforations or insertion-related infections. Complete expulsion occurred in 19 (7.9%) in the inserter and 13 (5.4%) forceps groups (p=.28). The inserter group had more partial expulsions (n=26, 10.8% versus n=12, 5.0%,) and self-removals (n=4, 1.7% versus n=0); however, most retained their IUD [174 (72.2%) and 200 (83.7%), respectively] (p=.01). Strings were not visible at follow-up in 23 (13.1%) participants in the inserter group and 45 (22.4%) in the forceps group (p=.04).
Conclusions:
High fundal placement was similar between groups, with strings subsequently seen more frequently in the inserter group. The inserter's increased partial expulsions and self-removals are likely associated with the longer string.
Implications:
The easily performed technique of Cu-T380A PPIUD insertion with the inserter may make immediate PPIUD insertions more convenient for patients and providers. The technique, involving less IUD manipulation and reduced dependence on extra instruments, could combine to make PPIUD a more popular option. Since this study has a short follow-up period, further study is needed to better understand comparative outcomes with longer follow-up. Additional research on string management for IUDs inserted with the inserter could help explain the increased number of partial expulsions.
More Related Videos
08:36Author Spotlight: Evaluating the Adjuvant Efficacy and Safety of Angong Niuhuang Pill in Viral Encephalitis Treatment
Published on: April 19, 2024
04:53A Clinical Trial Assessing the Safety, Efficacy, and Delivery of Olive-Oil-Based Three-Chamber Bags for Parenteral Nutrition
Published on: September 20, 2019
Related Concept Videos
Porin Insertion in the Outer Mitochondrial Membrane
Three models describe the assembly of porins by the SAM complex and their insertion into the outer membrane. Model 1 suggests that porins are assembled outside the SAM channel as the...
Insertion of Single-pass Transmembrane Proteins in the RER
Integral transmembrane proteins possess transmembrane and extra membrane domains. The transmembrane domains are primarily made of 20-25 hydrophobic amino acids arranged in a helical secondary confirmation. These...
Insertion of Multi-pass Transmembrane Proteins in the RER
The multipass transmembrane proteins are the type IV integral membrane proteins with multiple topogenic sequences determining their spatial arrangement in the ER membrane. Nearly all multipass proteins lack a cleavable signal sequence and use...
Survey Safety
Trial and Error and Algorithm
Clinical Trials: Overview