Quality of plasma samples and BD Vacutainer Barricor tubes: Effects of centrifugation
Andrea Padoan1, Martina Zaninotto2, Elisa Piva2
1Department of Medicine - DIMED, University of Padova, via Giustiniani 2, 35128 Padova, Italy; Department of Laboratory Medicine, University-Hospital of Padova, via Giustiniani 2, 35128 Padova, Italy.
Backgrounds:
The BD Vacutainer® Barricor™ Plasma collection tube (BD Barricor) uses an innovative non-gel separation method. This study compared the plasma residual cell count (PRCC) obtained from BD Barricor and from BD PST II plasma tubes.
Methods:
Four BD Barricors and one BD PST II were collected from 40 donors. BD PST II was centrifuged at 1300g/10 min, while the BD Barricors were centrifuged at 1800g/10 min, 4000g/3 min, 4000g/7 min and 4000g/15 min. PRCC was evaluated measuring white blood cells (WBC), red blood cells (RBC) and Platelets (PLT) counts by Siemens ADVIA 2120. Cell-free hemoglobin was quantified by haemolysis index (HI) by Roche Cobas c501.
Results:
BD PST II Median WBC, RBC and PLT counts were 0.38 (109/L), 0.0291 (1012/L) and 113.5 (109/L), respectively. Considering the BD PST II as reference, PRCC differences were expressed as median bias percentage. WBC showed a significant reduction at all the conditions (p < 0.01), being the reductions: 63.9% (1800g/10 min), 69.9% (4000g/3 min), 75.0% (4000g/7 min) and 82.7% (4000g/15 min). RBC reductions 29.7% (1800g/10 min), 33.8% (4000g/3 min), 39.6% (4000g/7 min) and 66.4 (4000g/15 min) were all significant (p < 0.01). PLT reductions were 1.6% at 1800g/10 min (p = ns), 1.2% at 4000g/3 min (p = ns), 27.1% at 4000g/7 min (p = 0.046) and 46.6% at 4000g/15 min (p = 0.005). BD Barricor centrifuged for 7 and 15 min at 4000g showed an increased haemolysis.
Conclusions:
BD Barricors plasma quality improved with increasing the centrifugation times but already at 4000g/3 min, the suggested centrifugation condition, a significant improvement was achieved.
More Related Videos
03:28Sperm Collection of Differential Quality Using Density Gradient Centrifugation
Published on: November 29, 2018
10:03Investigating the Detrimental Effects of Low Pressure Plasma Sterilization on the Survival of Bacillus subtilis Spores Using Live Cell Microscopy
Published on: November 30, 2017
Related Concept Videos
Centrifugation
Centrifugal Force
Uterine Tubes
Quality Control
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...
Quality Assurance
Quality of Water