Accuracy of device landing zone calcium volume measurement with contrast-enhanced multidetector computed tomography
Won-Keun Kim1, Matthias Renker2, Andreas Rolf3
1Kerckhoff Heart Center, Department of Cardiology, Bad Nauheim, Germany; DZHK (German Centre for Cardiovascular Research), partner site Rhein-Main, Frankfurt am Main, Germany; Kerckhoff Heart Center, Department of Cardiac Surgery, Bad Nauheim, Germany; Justus-Liebig University of Giessen, Department of Cardiology, Giessen, Germany.
Background:
The extent of aortic valve calcification is an important determinant of procedural success in transcatheter aortic valve implantation (TAVI). We sought to validate device landing zone calcium volume (DLZ-CV) measurements on contrast-enhanced multidetector computed tomography (MDCT) with non-contrast-enhanced scans as reference.
Methods:
We determined DLZ-CV in 141 patients undergoing transfemoral TAVI. Non-contrast-enhanced images were analyzed using a threshold of 130 HU as reference (DLZ-CV130). For contrast-enhanced scans, we applied various thresholds including 450 HU (DLZ-CV450), 850 HU (DLZ-CV850), mean aortic attenuation (AttenAo) + 2 SD (DLZ-CV2SD), AttenAo + 4 SD (DLZ-CV4SD), AttenAo + 4 SD + 5 mm3 volume filter (DLZ-CV4SD+), and based on visual estimation (DLZ-CVvis). We compared DLZ-CV values between patients with versus without paravalvular leak (PVL), and between patients with versus without post-dilatation stratified by the type of prosthesis.
Results:
All DLZ-CV measurements on contrast-enhanced scans significantly differed from DLZ-CV130 (p < 0.001 for all comparisons). The best approximation to DLZ-CV130 was achieved with DLZ-CV4SD+ (508 mm3 [332-772]; Pearson correlation: R = 0.87, p < 0.001; Bland-Altman: mean difference 1339 mm3 [limits of agreement 79;2600]). Moreover, DLZ-CV4SD+ allowed for discrimination of PVL ≥1° or the need for post-dilatation in patients receiving self-expanding prostheses. Procedural outcome using balloon-expandable prostheses was independent of DLZ-CV.
Conclusion:
Measurement of DLZ-CV using contrast-enhanced scans with unadjusted thresholds results in incorrect estimation of the calcium volume. The use of a scan-specific individual HU threshold including a volume filter (DLZ-CV4SD+) provides the best approximation to the reference and allows for discrimination of PVL ≥ 1° in patients receiving the Acurate neo prosthesis.
More Related Videos
13:09Segmentation and Measurement of Fat Volumes in Murine Obesity Models Using X-ray Computed Tomography
Published on: April 4, 2012
09:37Extracting Metrics for Three-dimensional Root Systems: Volume and Surface Analysis from In-soil X-ray Computed Tomography Data
Published on: April 26, 2016
Related Concept Videos
The Colonization of Land
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Uncertainty in Measurement: Accuracy and Precision
Imaging Studies III: Computed Tomography
Hybrid Zones
Improving Translational Accuracy