Optimization of Transesophageal Atrial Pacing to Assess Atrial Fibrillation Susceptibility in Mice
Published on: June 29, 2022
Atrioventricular Interval Extension Is Highly Efficient in Preventing Unnecessary Right Ventricular Pacing in
Valeria Calvi1, Ennio C Pisanò2, Marco Brieda3
1A.O.U. Policlinico Vittorio Emanuele, PO Ferrarotto, Catania, Italy.
Objectives:
This study sought to compare the Intrinsic Rhythm Support (IRSplus) and Ventricular Pace Suppress (VpS) in terms of right ventricular pacing percentage (VP %), mean atrioventricular interval (MAVI), atrial fibrillation, and cardiac volumes.
Background:
Modern pacemakers are provided with algorithms for reducing unnecessary ventricular pacing. These may be classified as: periodic search for intrinsic atrioventricular (AV) conduction prolonging the AV delay accordingly; or DDD-ADI mode switch. The IRSplus and VpS algorithms belong to the former and latter classes, respectively.
Methods:
Patients with sick sinus dysfunction without evidence of II/III degree AV block were 1:1 randomized to 6-month periods of either IRSplus or VpS, and then crossed over. Subsequent follow-ups were at the 12th month after randomization for device data retrieving, and at the 18th month with the same device programming for echocardiographic assessment.
Results:
A total of 230 patients (62% males, median age 75 years [interquartile range: 69 to 79 years]) were enrolled. At a linear mixed-model analysis with order of treatment and investigational sites as nested random effects, differences in VP% and MAVI reached statistical significance: VP% was 1% (0% to 11%) during IRSplus and 3% (0% to 26%) during VpS (p = 0.029); MAVI was 225 ms (198 to 253 ms) during IRSplus and 214 ms (188 to 240 ms) during VpS (p = 0.014). No differences were observed in atrial fibrillation burden and incidence, ejection fraction, and cardiac volumes.
Conclusions:
Both IRSplus and VpS algorithms ensured VP% ≤3% in most patients with sinus node dysfunction and preserved AV conduction. The IRSplus was slightly more efficient in reducing VP% at the expense of a small MAVI increase, with statistical but clinically insignificant differences. (Ventricular Pace Suppression Versus Intrinsic Rhythm Support Study; NCT01528657).
Related Concept Videos
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Coronary Artery Disease IV: Preventive Measures
Bioequivalence Experimental Study Designs: Completely Randomized and Randomized Block Designs
Prediction Intervals
However, the point estimate is most likely not the exact value of the population parameter, but close to it. After calculating point estimates, we construct interval estimates, called confidence intervals or prediction intervals. This prediction interval comprises a range of values unlike the point estimate and is a better predictor of the observed sample value, y.
Switching of BJT
Cut-off Mode ("Off" State): In this state, both the emitter-base and collector-base junctions are...
Crossing Over
The homologous pairs of sister chromosomes—one from the maternal and one from the paternal genome—then begin to align alongside each other lengthwise, matching corresponding DNA positions in a process...