Novel Extravascular Defibrillation Configuration With a Coil in the Substernal Space: The ASD Clinical Study
Joseph Y S Chan1, Jacek Lelakowski2, Francis D Murgatroyd3
1Division of Cardiology, Department of Medicine and Therapeutics, Prince of Wales Hospital, Hong Kong, The Chinese University of Hong Kong, Hong Kong.
Objectives:
This study assessed the defibrillation efficacy of the substernal-lateral electrode configuration.
Background:
Subcutaneous implantable cardioverter-defibrillators (ICDs) are regarded as alternatives to transvenous ICDs in certain subjects. However, substantially higher shock energy of up to 80 J may be required. Proposed is a new defibrillation method of placing the shock coil into the substernal space.
Methods:
This prospective, nonrandomized, feasibility study was conducted in subjects scheduled for midline sternotomy or implant of ICD. A blunted end tunneling tool was used to insert a defibrillation lead behind the sternum using a percutaneous subxiphoid approach. A skin patch electrode was placed on the left mid-axillary line at the fourth to fifth intercostal space. After ventricular fibrillation induction, a single 35-J shock was delivered between the lead and skin patch.
Results:
Sixteen subjects (12 males, 4 females; mean age: 61.6 ± 11.8 years) were enrolled. The mean lead placement time was 11.1 ± 6.6 min. Of the 14 subjects with successfully induced ventricular fibrillation episodes, 13 subjects (92.9%) had successful defibrillation. The 1 failure was associated with high and lateral shock coil placement. Mean ventricular fibrillation duration was 18.4 ± 5.6 s with a shock impedance of 98.1 ± 19.3 ohms. Of the 11 subjects with coil-patch electrograms, the average R-wave amplitude during sinus rhythm was 3.0 ± 1.4 mV.
Conclusions:
These preliminary data demonstrate that substernal defibrillation is feasible and successful defibrillation can be achieved with the shock energy available in current transvenous ICDs. This may open new alternatives to extravascular ICD therapy.
More Related Videos
Related Concept Videos
Controller Configurations
Control-system compensation involves various configurations, most commonly series or cascade compensation, in which the controller...
Electron Configurations
The relative energies of the subshells determine the order in which atomic orbitals are filled (1s, 2s, 2p, 3s, 3p,...
Electron Configuration of Multielectron Atoms
Two-Compartment Open Model: Extravascular Administration
The absorption exponent (ka) indicates the speed at which the drug...
Configurations of BJT
The common base configuration is noted for its high voltage gain, positioning it as an ideal choice for single-stage amplifier circuits, such as microphone pre-amplifiers. A notable...
Stability of Equilibrium Configuration
A stable equilibrium occurs when a system tends to return to its original position when given a small displacement, and the potential energy is at its minimum. An example of a stable equilibrium is when a cantilever beam is fixed at one end and a weight is attached to the other end. If the weight...