Subcutaneous Versus Transvenous Implantable Defibrillator Therapy: A Meta-Analysis of Case-Control Studies
Indranill Basu-Ray1, Jing Liu2, Xiaoming Jia2
1Department of Cardiovascular Disease, Texas Heart Institute, Houston, Texas.
Objectives:
This study aims to conduct a meta-analysis comparing efficacy and safety outcomes between subcutaneous implantable cardioverter-defibrillator (S-ICD) and transvenous implantable cardioverter-defibrillator (TV-ICD).
Background:
The S-ICD was developed to minimize complications related to the conventional TV-ICD. Direct comparison of clinical outcomes between the 2 devices has been limited by varying patient characteristics and definitions of complications with no randomized trials completed comparing these systems.
Methods:
Studies in the PubMed and Embase databases and secondary referencing sources were systematically reviewed. Studies meeting criteria were included in the meta-analysis. Baseline characteristics and outcome data of the S-ICD and TV-ICD groups were appraised and analyzed. A random-effects model was used to derive odds ratio (OR) with 95% confidence interval (CI).
Results:
Five studies met inclusion criteria. Baseline characteristics were similar between the S-ICD and TV-ICD groups. Fewer lead complications occurred in the S-ICD group compared to the TV-ICD group (OR: 0.13; 95% CI: 0.05 to 0.38). The infection rate was similar between the S-ICD and TV-ICD groups (OR: 0.75; 95% CI: 0.30 to 1.89). There were no differences in system or device failures between groups (OR: 1.13; 95% CI: 0.43 to 3.02). Overall, inappropriate therapy (T-wave oversensing, supraventricular tachycardia, episodes of inappropriate sensing) was similar between the 2 groups (OR: 0.87; 95% CI: 0.51 to 1.49). However, the nature of inappropriate therapy was different between the S-ICD and TV-ICD groups. Both devices appear to perform equally well with respect to appropriate shocks.
Conclusions:
S-ICD reduced lead-related complications but was similar to TV-ICD with regard to non-lead-related complications, including inappropriate therapy. These results support the concept that S-ICD is a safe and effective alternative to TV-ICD in appropriate patients.
More Related Videos
06:26Meta-analysis of Voxel-Based Neuroimaging Studies using Seed-based d Mapping with Permutation of Subject Images SDM-PSI
Published on: November 27, 2019
08:36Author Spotlight: Evaluating the Adjuvant Efficacy and Safety of Angong Niuhuang Pill in Viral Encephalitis Treatment
Published on: April 19, 2024
Related Concept Videos
Types of Biopharmaceutical Studies: Controlled and Non-Controlled Approaches
Non-controlled studies, commonly employed for initial exploration, lack a control group, rendering them susceptible to biases and external influences. In contrast,...
Directing Effect of Substituents: meta-Directing Groups
Gene Therapy
Group Therapy
Inflammatory Bowel Disease III: Diagnostic Studies and Management I-Nutritional Therapy
Diagnostic studies
A colonoscopy is the definitive screening test, distinguishing ulcerative colitis from other colon diseases with similar symptoms. During a colonoscopy test, inflamed mucosa with exudate ulcerations can be observed, and biopsies are taken to determine the histologic characteristics of the...
meta-Directing Deactivators: –NO2, –CN, –CHO, –⁠CO2R, –COR, –CO2H