A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Acute and 3-Month Performance of a Communicating Leadless Antitachycardia Pacemaker and Subcutaneous Implantable
Fleur V Y Tjong1, Tom F Brouwer1, Brendan Koop2
1Academic Medical Center, Amsterdam, the Netherlands.
Objectives:
The primary objective was to assess the acute and 3-month performance of the modular antitachycardia pacing (ATP)-enabled leadless pacemaker (LP) and subcutaneous implantable cardioverter-defibrillator (S-ICD) system, particularly device-device communication and ATP delivery.
Background:
Transvenous pacemakers and implantable cardioverter-defibrillators (ICDs) have considerable rates of lead complications. We examined the next step in multicomponent leadless cardiac rhythm management: feasibility of pacing (including ATP) by a LP, commanded by an implanted S-ICD through wireless, intrabody, device-device communication.
Methods:
The combined modular cardiac rhythm management therapy system of the LP and S-ICD prototypes was evaluated in 3 animal models (ovine, porcine, and canine) both in acute and chronic (90 days) experiments. LP performance, S-ICD to LP communication, S-ICD and LP rhythm discrimination, and ATP delivery triggered by the S-ICD were tested.
Results:
The LP and S-ICD were successfully implanted in 98% of the animals (39 of 40). Of the 39 animals, 23 were followed up for 90 days post-implant. LP performance was adequate and exhibited appropriate VVI behavior during the 90 days of follow-up in all tested animals. Unidirectional communication between the S-ICD and LP was successful in 99% (398 of 401) of attempts, resulting in 100% ATP delivery by the LP (10 beats at 81% of the coupling interval). Adequate S-ICD sensing was observed during normal sinus rhythm, LP pacing, and ventricular tachycardia/ventricular fibrillation.
Conclusions:
This study presents the preclinical acute and chronic performance of the combined function of an ATP-enabled LP and S-ICD. Appropriate VVI functionality, successful wireless device-device communication, and ATP delivery were demonstrated by the LP. Clinical studies on safety and performance are needed.
Related Concept Videos
Communication
Communication
Neuronal Communication
Therapeutic Communication
Verbal communication depends on language or a prescribed way of using words so that people can share information effectively. The critical aspects of verbal...
Barriers to Effective Communication II
Cultural barriers:
Differences in values, beliefs, religion, knowledge, and tradition can significantly impact communication. Awareness of nonverbal cues is critical, especially when conversing with a patient from a different culture. What appears appropriate in one culture may be inappropriate in another.
Semantic barriers:
As a result of their tendency to use...
Channels of Non-Verbal Communication