Comparison of image quality and radiation exposure between conventional imaging and gemstone spectral imaging in
Tianqi Fang1, Wei Deng2,3, Martin Wai-Ming Law4
11 National-Regional Key Technology Engineering Laboratory for Medical Ultrasound, Guangdong Key Laboratory for Biomedical Measurements and Ultrasound Imaging, School of Biomedical Engineering, Health Science Centre, Shenzhen University , Shenzhen , China.
Objective:
To compare patients' image quality and radiation exposure between gemstone spectral imaging (GSI) with rapid kV switching technique and conventional polychromatic imaging (CPI) performed in abdominal CT examinations.
Methods:
Adult patients who were referred to abdominal CT from October 2015 to March 2016 were enrolled. Unenhanced CT with CPI mode and tri-phase (arterial/portal/delayed phase) contrast-enhanced scan with GSI mode were performed with different protocols respectively. Regions of interest (ROIs) were drawn on muscle and fat. Parametric results of the image noise, signal-to-noise ratio (SNR) and clinical image quality in these regions between the monochromatic images reconstructed at 65 keV and conventional polychromatic images were compared. Radiation dose was also compared between CPI and GSI.
Results:
43 patients were recruited. Compared to conventional imaging, the noise level was generally not significantly different between GSI images in arterial phase and portal phase, and significantly higher (around 10%) in delayed phase. The SNR of GSI in portal phase was significantly higher than that of conventional imaging, and was similar between arterial phase/delayed phase and conventional imaging. The clinical image quality between conventional imaging and GSI was generally not significantly different. The dose length product was reduced by 0.3-20.1% in GSI compared to conventional imaging.
Conclusion:
GSI reduces the radiation exposure slightly, however maintains or even improves image quality. These results may warrant the application of GSI in patients referred for abdominal CT. Advances in knowledge: Compared to CPI, GSI reduces the radiation exposure slightly, however maintains or even improves image quality in abdominal CT. These findings may warrant the application of GSI in patients referred for abdominal CT.
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies VII: Vascular Imaging
X-ray Imaging
Brain Imaging
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic...