Related Experiment Video
Updated: Feb 10, 2026

Complete and Partial Resuscitative Endovascular Balloon Occlusion of the Aorta for Hemorrhagic Shock
Published on: May 19, 2022
Resuscitative endovascular balloon occlusion of the aorta: rupture risk and implications for blind inflation
Philip J Wasicek1, William A Teeter1, Megan L Brenner1
1Program in Trauma/Critical Care, R Adam Cowley Shock Trauma Center, University of Maryland Medical System, Baltimore, Maryland, USA.
Background:
Resuscitative endovascular balloon occlusion of the aorta (REBOA) is a torso hemorrhage control technique. To expedite deployment, inflation is frequently performed as a blind technique with minimal imaging, which carries a theoretical risk of aortic injury. The objective of this study was to examine the relationship between balloon inflation, deformation and the risk of aortic rupture.
Methods:
Compliant balloon catheters were incrementally inflated in segments of cadaveric swine aorta. Serial longitudinal and circumferential measurements were recorded, along with the incidence of aortic rupture.
Results:
Fourteen cadaveric swine aorta segments were tested with mean (±SD) baseline aortic diameter (mm) of 14.2±3.4. Rupture occurred in three aortas. The mean baseline diameters (mm) of the aortic segments that were ruptured were significantly smaller than those that did not rupture (8.9±1.2 vs 15.6±1.9; P<0.001). The maximal circumferential stretch ratios were significantly higher in the aorta segments that ruptured compared with those that did not (1.9±0.1 vs 1.5±0.1; P<0.001). The maximal amount of balloon longitudinal deformation was 80 mm (116% longer than the intended working length).
Conclusions:
Inflation of aortic balloon catheters carries an inherent risk of aortic injury, which may be minimized through an understanding of the intrinsic characteristics of the aorta and compliant balloons. Smaller diameter aortic segments undergoing overinflation, particularly beyond a circumferential stretch ratio of 1.8, are at risk of aortic rupture.
Level Of Evidence:
Level II.