Knee dGEMRIC at 7 T: comparison against 1.5 T and evaluation of T1-mapping methods
Pernilla Peterson1,2, Carl Johan Tiderius3, Emma Olsson4
1Medical Radiation Physics, Department of Translational Medicine, Lund University, Inga Marie Nilssons gata 49, SE-205 02, Malmö, Sweden. pernilla.peterson@med.lu.se.
Background:
dGEMRIC (delayed Gadolinium Enhanced Magnetic Resonance Image of Cartilage) is a well-established technique for cartilage quality assessment in osteoarthritis at clinical field strengths. The method is robust, but requires injection of contrast agent and a cumbersome examination procedure. New non-contrast-agent-based techniques for cartilage quality assessment are currently being developed at 7 T. However, dGEMRIC remains an important reference technique during this development. The aim of this work was to compare T1 mapping for dGEMRIC at 7 T and 1.5 T, and to evaluate three T1-mapping methods at 7 T.
Methods:
The knee of 10 healthy volunteers and 9 patients with early signs of cartilage degradation were examined at 1.5 T and 7 T after a single (one) contrast agent injection (Gd-(DTPA)2-). Inversion recovery (IR) sequences were acquired at both field strengths, and at 7 T variable flip angle (VFA) and Look-Locker (LL) sequences were additionally acquired. T1 maps were calculated and average T1 values were estimated within superficial and deep regions-of-interest (ROIs) in the lateral and medial condyles, respectively.
Results:
T1 values were 1.8 (1.4-2.3) times longer at 7 T. A strong correlation was detected between 1.5 T and 7 T T1 values (r = 0.80). For IR, an additional inversion time was required to avoid underestimation (bias±limits of agreement - 127 ± 234 ms) due to the longer T1 values at 7 T. Out of the two 3D sequences tested, LL resulted in more accurate and precise T1 estimation compared to VFA (average bias±limits of agreement LL: 12 ± 202 ms compared to VFA: 25 ± 622 ms). For both, B1 correction improved agreement to IR.
Conclusion:
With an adapted sampling scheme, dGEMRIC T1 mapping is feasible at 7 T and correlates well to 1.5 T. If 3D is to be used for T1 mapping of the knee at 7 T, LL is preferred and VFA is not recommended. For VFA and LL, B1 correction is necessary for accurate T1 estimation.
Related Concept Videos
Knee Joint
A total of seven ligaments support the knee joint. The patellar ligament, which is also attached to the quadriceps femoris...
Renal Drug Clearance: Comparison Between Renal Excretion Methods
Renal clearance is often associated with the renal glomerular filtration rate (GFR), which represents the rate at which plasma is filtered through the glomeruli in the kidney. When drug reabsorption is minimal and there is no active secretion, renal clearance is closely related to the...
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Evolutionary Relationships through Genome Comparisons
Comparison Between Electrical And Gravitational Forces
Since both are inverse square law forces, the distance gets canceled when the ratio of the two forces is considered. Instead, the ratio of the electrical and gravitational forces depends on...