Mycobacterium chimaera infections in post-operative patients exposed to heater-cooler devices: An overview
T Ogunremi1, G Taylor2, L Johnston3
1Centre for Communicable Diseases and Infection Control, Public Health Agency of Canada, Ottawa, ON.
Abstract:
A multi-country outbreak of Mycobacterium chimaera infection associated with contaminated heater-cooler devices (HCDs) has been reported, with more than 70 cases in Europe and the United States and two cases in Canada to date. The epidemiological and microbiological characteristics of this outbreak provide evidence for common-source transmission of M. chimaera from the exhaust air of intrinsically contaminated HCDs to patients during cardiac surgery. To date, all reported cases have been associated with Stöckert 3T HCDs manufactured at one plant by LivaNova prior to September 2014. Implantation of prosthetic material increases the risk of infection. Infections usually present as prosthetic valve endocarditis, vascular graft infection or disseminated infection. Reported mortality rates have varied, but were often over 40%. Several measures are recommended to facilitate case-finding and mitigate risk of exposure. The feasibility of some risk mitigation measures and their effectiveness in reducing the risk of exposure are yet to be determined. Until HCDs are redesigned in a manner that prevents water contamination and aerosolization, separating the HCD exhaust air from the operating room air during surgery may be the most effective risk mitigation strategy. However, possible unintended consequences of this approach should be considered. This overview summarizes findings from peer-reviewed and other relevant national documents on key features of the outbreak, including the source, identified risk factors for infection, signs and symptoms of infection, burden of disease, risk mitigation measures, management challenges and knowledge gaps.
Insights
A multi-country outbreak of Mycobacterium chimaera infections linked to heater-cooler devices (HCDs) has occurred. Separating HCD exhaust from operating rooms may mitigate transmission during cardiac surgery.
Area of Science:
- Infectious Diseases
- Public Health
- Medical Device Safety
Background:
- A significant multi-country outbreak of Mycobacterium chimaera infections has been identified.
- These infections are linked to contaminated heater-cooler devices (HCDs) used during cardiac surgery.
- Over 70 cases have been reported in Europe and the United States, with two in Canada.
Purpose of the Study:
- To summarize key features of the Mycobacterium chimaera outbreak.
- To detail the source, risk factors, clinical presentation, and disease burden.
- To review recommended risk mitigation strategies and identify knowledge gaps.
Main Methods:
- Review of peer-reviewed literature and national documents.
- Epidemiological and microbiological analysis of outbreak data.
- Assessment of risk factors and clinical characteristics of affected patients.
Main Results:
- Common-source transmission of M. chimaera from HCD exhaust air to patients during surgery is evidenced.
- All cases are associated with specific Stöckert 3T HCDs manufactured before September 2014.
- Infections typically manifest as prosthetic valve endocarditis or disseminated disease, with high mortality rates (>40%).
Conclusions:
- Separating HCD exhaust from operating room air is a potential risk mitigation strategy.
- Redesigning HCDs to prevent water contamination is crucial.
- Further research is needed to determine the feasibility and effectiveness of mitigation measures.
Related Concept Videos
Cardiomyopathy VII: Pre and Post Operative Nursing Management
Pharmacokinetics in Pediatric Patients: Overview and Drug Absorption
Real Number Operations
Operational Amplifiers
Vector Operations
A vector multiplied by a scalar value is called scalar multiplication. The result obtained is a new vector with a different magnitude. If the scalar is positive, the direction of the vector remains the same, but if it is negative, the direction of the vector is reversed. For example, the product of the mass and velocity yields the momentum.
Operant Conditioning
Reinforcement in operant conditioning can be positive or negative, both of which serve to increase the likelihood of a behavior. Positive...